Evaluation of a commercial radiometric system for primary isolation of mycobacteria over a fifteen-year period.

作者信息

Middleton A M, Chadwick M V, Gaya H

机构信息

Department of Microbiology, Royal Brompton Hospital, London, UK.

出版信息

Eur J Clin Microbiol Infect Dis. 1997 Feb;16(2):166-70. doi: 10.1007/BF01709480.

Abstract

The rate of recovery and time to detection of mycobacteria from clinical specimens were analysed for specimens inoculated into both radiometric Middlebrook 7H12 (Bactec 12B, originally 12A) medium and Lowenstein-Jensen egg-based medium (with and without sodium pyruvate) over the 15-year period from 1980 to 1994. A total of 19,679 Bactec vials were inoculated together with Lowenstein-Jensen slopes, with 2198 mycobacterial isolates detected. The mean times to detection for Bactec and Lowenstein-Jensen were 11.7 days and 23.9 days, respectively. In 195 cases mycobacteria were isolated from Bactec and not from Lowenstein-Jensen, whereas the reverse was true in 42 cases. The number of contaminated Bactec vials was 488 and the number of contaminated Lowenstein-Jensen slopes 448.

摘要

对1980年至1994年这15年间接种于放射性米氏7H12培养基(Bactec 12B,原为12A)和罗氏鸡蛋培养基(含和不含丙酮酸钠)的临床标本中分枝杆菌的恢复率和检测时间进行了分析。共接种了19679个Bactec瓶和罗氏斜面,检测到2198株分枝杆菌分离株。Bactec和罗氏培养基的平均检测时间分别为11.7天和23.9天。195例中,分枝杆菌从Bactec分离而未从罗氏培养基分离,反之,42例则相反。污染的Bactec瓶数量为488个,污染的罗氏斜面数量为448个。