Yang Y C, Hulbert M F, Batterbury M, Clearkin L G
St. Paul's Eye Unit, Liverpool, England.
J Glaucoma. 1997 Jun;6(3):175-9.
To determine the reproducibility and the normal reference range of pulsatile ocular blood flow (POBF) values in healthy subjects using the Ocular Blood Flow Tonograph (OBF Laboratories, UK Ltd., Wiltshire, England).
Pulsatile ocular blood flow was measured in one eye of each of 83 patients. Coefficient of reliability was determined by calculation of intraclass correlation coefficient via one-way analysis of variance. Mean difference between measurements was calculated for bias and first exposure effects. Pulsatile ocular blood flow from 163 healthy individuals were analyzed to determine the distribution, mean, standard deviation (SD), range, and the 5th and 95th percentile values. The influence of age, blood pressure, pulse rate, and intraocular pressure on pulsatile ocular blood flow was determined by regression analysis.
Reliability coefficient for pulsatile ocular blood flow values ranging from 290 microliters/min to 2,196 microliters/min was 0.92. Variation in bias and first exposure effect were not significant. Pulsatile ocular blood flow values were normally distributed. Mean values were 669.90 +/- 233.0 microliters/min in men and 841.90 +/- 254.6 microliters/min in women. Fifth and ninety-fifth percentile values were 364.75 microliters/min and 1,266.10 microliters/min in men and 397.18 microliters/min and 1,346.10 microliters/min in women. Pulsatile ocular blood flow was significantly influenced by pulse rate.
This study confirms the reliability of the Ocular Blood Flow Tonograph in repeated measurements of POBF within individuals over short time intervals. The high interindividual variation in POBF may invalidate comparison of POBF between individuals, and the wide range of normal values may limit the value of using a low POBF as a possible indicator of disease.
使用眼血流眼压描记仪(英国威尔特郡OBF实验室有限公司)测定健康受试者搏动性眼血流量(POBF)值的可重复性和正常参考范围。
对83例患者的单眼进行搏动性眼血流量测量。通过单向方差分析计算组内相关系数来确定可靠性系数。计算测量值之间的平均差异以评估偏差和首次暴露效应。分析163名健康个体的搏动性眼血流量,以确定其分布、均值、标准差(SD)、范围以及第5和第95百分位数。通过回归分析确定年龄、血压、脉搏率和眼压对搏动性眼血流量的影响。
搏动性眼血流量值在290微升/分钟至2196微升/分钟之间的可靠性系数为0.92。偏差和首次暴露效应的变化不显著。搏动性眼血流量值呈正态分布。男性的平均值为669.90±233.0微升/分钟,女性为841.90±254.6微升/分钟。男性的第5和第95百分位数分别为364.75微升/分钟和1266.10微升/分钟,女性为397.18微升/分钟和1346.10微升/分钟。搏动性眼血流量受脉搏率的显著影响。
本研究证实了眼血流眼压描记仪在短时间内对个体进行POBF重复测量的可靠性。POBF个体间差异较大可能使个体间POBF的比较无效,且正常范围较宽可能限制将低POBF用作疾病可能指标的价值。