Performance evaluation of a diode array for enhanced dynamic wedge dosimetry.

作者信息

Zhu T C, Ding L, Liu C R, Palta J R, Simon W E, Shi J

机构信息

Department of Radiation Oncology, University of Florida, Gainesville 32610-0385, USA.

出版信息

Med Phys. 1997 Jul;24(7):1173-80. doi: 10.1118/1.598019.

Abstract

The performance of a diode array (Profiler) was evaluated by comparing its enhanced dynamic wedge (EDW) profiles measured at various depths with point measurements using a 0.03 cm3 ionization chamber on a commercial linear accelerator. The Profiler, which covers a 22.5 cm width, was used to measure larger field widths by concatenating three data sets into a larger field. An innovative wide-field calibration technique developed by the manufacturer of the device was used to calibrate the individual diode sensitivity, which can vary by more than 10%. Profiles of EDW measured with this device at several depths were used to construct isodose curves using the percentage depth dose curve measured by the ionization chamber. These isodose curves were used to check those generated by a commercial treatment planning system. The profiles measured with the diode array for both 8 and 18 MV photon beams agreed with those of the ionization chamber within a standard deviation of 0.4% in the field (defined as 80% of the field width) and within a maximum shift of less than 2 mm in the penumbra region. The percentage depth dose generally agreed to within 2% except in the buildup region. The Profiler was extremely useful as a quality assurance tool for EDW and as a dosimetry measurement device with tremendous savings in data acquisition time.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验