[Quantification of dehydration of soda lime in clinical circumstances].

作者信息

Soro-Domingo M, Cortés-Uribe A, Alvarez-Refojo F, Bonome C, Belda Nacher F J

机构信息

Servicio de Anestesiología, Hospital Virgen de la Salud, Toledo.

出版信息

Rev Esp Anestesiol Reanim. 1997 May;44(5):186-90.

PMID:9280996
Abstract

HYPOTHESIS AND OBJECTIVES

When the CO2 absorbents, soda lime and baralime, have lost their normal level of hydration, they may react with certain halogenated anesthetics to produce appreciable levels of carbon monoxide. The degree of absorbent desiccation has been considered the limiting factor for this phenomenon. This study quantifies the level of dehydration of lime produced under clinical conditions and the influence of several factors.

MATERIAL AND METHOD

Desiccation was determined: 1) at set periods of time (3, 7 and 14 days) after clinical use of fresh soda lime in general anesthesia using a fresh gas flow (FGF) of 6 l/min, and 2) after gas had been crossing the continuous flow (CF) oxygen reservoir at 7 l/min for 17 and 65 hours. Two anesthetic systems were used: a) the Ohmeda Excel-210, in which the continuous FGF did not cross the reservoir and b) the Siemens Ventilator 710, in which the FGF did cross the reservoir. The experiments were repeated with three types of lime.

RESULTS

The clinical use of lime for 3, 7 and 14 days caused different levels of desiccation, with decreases in hydration of up to 50% and 14 days. Nevertheless, water content was always over 5%, a level at which no reaction with halogenated agents takes place. After 17 and 65 hours of CF in the circuit where continuous FGF did not pass through the canister, the water content did not change. With the Siemens 710 circuit, in which the continuous FGF crossed the canister, the dehydration level was 1.2 +/- 0.3% after 17 hours and 0.7 +/- 0.3% after 65 hours, a level that can produce CO upon reaction between lime and halogenated gases. The type of lime used had little effect.

CONCLUSIONS

Lime does not desiccate to levels able to produce CO in daily use, regardless of the FGF system used. The phenomenon of desiccation depends on two factors: 1) use of anesthetic equipment in which continuous FGF conditions require gas to pass through the canister, and 2) the maintenance of CF for a sufficient period of time.