Minami K, Arusoglu L, Koyanagi T, el-Banayosy A, Körner M M, Körfer R
Department of Thoracic and Cardiovascular Surgery, University of Bochum, Bad Oeynhausen, Germany.
Ann Thorac Surg. 1997 Sep;64(3):861-2. doi: 10.1016/s0003-4975(97)00692-9.
Persistent oozing and bleeding through the wall of the built-in outflow conduit in mechanical circulatory assist devices is a troublesome problem. Wrapping the outflow conduit of Thoratec in a Hemashield graft without preclotting completely prevents oozing and bleeding.
在机械循环辅助装置中,内置流出管道壁持续渗血是一个棘手的问题。将Thoratec流出管道包裹在未预凝的Hemashield移植物中可完全防止渗血。