Jakob P M, Griswold M A, Lovblad K O, Chen Q, Edelman R R
Department of Radiology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts 02215, USA.
Magn Reson Med. 1997 Oct;38(4):534-40. doi: 10.1002/mrm.1910380405.
A modified BURST imaging technique with a nearly twofold reduction in acquisition time, improved signal-to-noise ratio and clear improvement of the image quality is introduced. The method applies half-Fourier reconstruction and can be implemented on a standard clinical scanner. This approach allows the collection of single-shot magnetization prepared images of the human brain with a matrix size of 112 x 128 within 86-250 ms. The signal-to-noise properties of the pulse sequence are analyzed and compared with a standard BURST experiment.