Foltz W D, Stainsby J A, Wright G A
Department of Medical Biophysics, University of Toronto, Ontario, Canada.
Magn Reson Med. 1997 Nov;38(5):759-68. doi: 10.1002/mrm.1910380512.
MR oximetry requires a T2 measurement that is accurate within 5% in vivo. Simple methods are susceptible to signal loss and tend to underestimate T2. Current methods utilize RF pulses or RF cycling patterns that prevent signal loss at each data acquisition. However, using these methods with imperfect pulses, T2 tends to be overestimated due to temporary storage of the magnetization along the longitudinal axis where it decays more slowly with a time constant T1 > T2. To reduce the T1 dependence while preventing signal loss, we utilize simple 90x180y90x composite pulses and good RF cycling patterns. These trains are critical for T2 accuracy over typical ranges of RF and static field inhomogeneities and refocusing intervals. T1 signal decay during each 90x180y90x pulse must be accounted for to yield accuracy within 5% when the pulse-width is 10% or more of the refocusing interval. A simple correction scheme compensates for this T1-related error effectively.
磁共振血氧测定法需要在体内实现5%以内的准确T2测量。简单方法易出现信号丢失,且往往会低估T2。当前方法利用射频脉冲或射频循环模式来防止每次数据采集时的信号丢失。然而,使用这些带有不完美脉冲的方法时,由于沿着纵向轴的磁化强度会暂时存储,其以时间常数T1>T2的速度衰减得更慢,T2往往会被高估。为了在防止信号丢失的同时减少对T1的依赖,我们采用简单的90x180y90x复合脉冲和良好的射频循环模式。这些序列对于在射频和静磁场不均匀性以及重聚焦间隔的典型范围内实现T2准确性至关重要。当脉冲宽度为重聚焦间隔的10%或更多时,必须考虑每个90x180y90x脉冲期间的T1信号衰减,以实现5%以内的准确性。一种简单的校正方案能有效补偿这种与T1相关的误差。