Auffarth G U, Tetz M R, Biazid Y, Völcker H E
Department of Ophthalmology, Ruprecht-Karls-University of Heidelberg, Germany.
J Cataract Refract Surg. 1997 Nov;23(9):1351-5. doi: 10.1016/s0886-3350(97)80114-9.
To assess the reliability of the Orbscan Topography System in measuring anterior chamber depth (ACD).
Department of Ophthalmology, Ruprecht-Karls-University of Heidelberg, Heidelberg, Germany.
The Orbscan Topography System is a 3-D, scanning slit beam system for analyzing corneal surfaces as well as structures of the anterior segment, e.g., iris, lens. Reflected surface data points are measured in the x, y, and z axes, creating color-coded true surface topography maps. We measured ACD prior to cataract surgery in 56 eyes of 37 patients (mean age 70.4 years + 13.2 [SD]) using the Orbscan system and, as a control, the Grieshaber Biometric System (ultrasound) and the Jäger ACD slitlamp measurement system.
Mean ACD values measured with the Orbscan system were 3.23 +/- 0.55 mm; ultrasound measurements were 3.27 +/- 0.54 mm and Jäger measurements, 3.10 +/- 0.55 mm. The average difference between Orbscan and ultrasound values was 0.04 +/- 0.15 mm. The correlation coefficient between Orbscan and ultrasound was 0.96 (P < .00001) and between Orbscan and Jäger, 0.899 (P < .0001).
The ACD measurements by the Orbscan system were equivalent to common reference measurements.
评估Orbscan地形测量系统测量前房深度(ACD)的可靠性。
德国海德堡鲁普雷希特-卡尔斯大学眼科。
Orbscan地形测量系统是一种三维扫描裂隙光束系统,用于分析角膜表面以及眼前节结构,如虹膜、晶状体。在x、y和z轴上测量反射表面数据点,生成彩色编码的真实表面地形图。我们使用Orbscan系统在37例患者(平均年龄70.4岁±13.2[标准差])的56只眼中测量白内障手术前的ACD,并作为对照,使用Grieshaber生物测量系统(超声)和Jäger ACD裂隙灯测量系统。
Orbscan系统测量的平均ACD值为3.23±0.55mm;超声测量值为3.27±0.54mm,Jäger测量值为3.10±0.55mm。Orbscan和超声值之间的平均差异为0.04±0.15mm。Orbscan与超声之间的相关系数为0.96(P<.00001),Orbscan与Jäger之间的相关系数为0.899(P<.0001)。
Orbscan系统测量的ACD与常用参考测量值相当。