Lee Jong Yun, Kim Jun-Heon, Kim Hyo-Myung, Song Jong-Suk
Dept of Ophthalmology, Korea University College of Medicine, 80 Guro-dong, Guro-gu, Seoul 152-703, Republic of Korea.
J Refract Surg. 2007 May;23(5):487-91. doi: 10.3928/1081-597X-20070501-12.
To compare the measurement of anterior chamber depth exclusive of corneal thickness using Orbscan IIz and ultrasound biomicroscopy (UBM) and evaluate the repeatability of each method.
Three consecutive measurements of anterior chamber depth were prospectively performed using Orbscan IIz and Paradigm 50-MHz UBM in 40 eyes in 20 individuals. Mean values were compared using the paired t test. For 12 eyes in 6 individuals, anterior chamber depth measurements were performed 5 times to estimate the repeatability of each method by a coefficient of variation. Refractive errors were measured to correlate with anterior chamber depth.
The mean anterior chamber depth was 2.82 +/- 0.46 mm with the Orbscan IIz and 2.91 +/- 0.43 mm in UBM. This difference was statistically significant (P < .001), but not clinically meaningful. The coefficient of variation was 1.15% and 1.10% in Orbscan IIz and UBM, respectively. A negative correlation between anterior chamber depth and spherical equivalent refraction was noted within the range of -5.50 to +3.00 diopters.
The mean anterior chamber depth of Orbscan IIz was 0.087 mm less than that of UBM. Both methods were precise. Orbscan IIz seems to be a useful and more convenient method to measure anterior chamber depth for phakic intraocular lens implantation.
比较使用Orbscan IIz和超声生物显微镜(UBM)测量不包括角膜厚度的前房深度,并评估每种方法的可重复性。
对20名个体的40只眼前瞻性地使用Orbscan IIz和Paradigm 50-MHz UBM连续测量前房深度3次。使用配对t检验比较平均值。对6名个体的12只眼进行5次前房深度测量,通过变异系数估计每种方法的可重复性。测量屈光不正以与前房深度相关联。
Orbscan IIz测量的平均前房深度为2.82±0.46mm,UBM测量的为2.91±0.43mm。这种差异具有统计学意义(P<.001),但无临床意义。Orbscan IIz和UBM的变异系数分别为1.15%和1.10%。在-5.50至+3.00屈光度范围内,前房深度与等效球镜度呈负相关。
Orbscan IIz测量的平均前房深度比UBM少0.087mm。两种方法都很精确。对于有晶状体眼人工晶状体植入,Orbscan IIz似乎是一种有用且更方便的测量前房深度的方法。