Jarvik R, Westaby S, Katsumata T, Pigott D, Evans R D
Oxford Heart Centre, John Radcliffe Hospital, United Kingdom.
Ann Thorac Surg. 1998 Feb;65(2):470-3. doi: 10.1016/s0003-4975(97)01343-x.
Driveline infection limits the event-free survival of patients with a left ventricular assist device. With the evolving prospect of improved left ventricular assist devices in the bridge-to-transplantation or recovery setting, we sought to reduce the risk of driveline complications.
As part of the Oxford Jarvik 2000 research program, we developed a carbon and then titanium pedestal to transmit the electric wires through the skin. In a sheep model, the pedestal was brought out through the skin of the shoulder (n = 10) or the scalp (n = 9) with underlying fixation to the skull. Exit wounds were carefully inspected for healing and infection. Power cable durability tests were performed in 6 additional animals without an implanted pump.
The cumulative observation period was 1,491 days (mean time, 78 days; range, 14 days to 198 days). There was no difference in observation period between the two groups. Infection (n = 2) and impaired healing (n = 5) occurred in the mobile tissues at the shoulder. Skull-mounted pedestals were free from infection or healing problems. The electric cables were not interrupted by repeated neck flexion (cumulative observation period, 588 days). The carbon pedestal was replaced by a titanium pedestal when the head butting of the sheep fractured the carbon.
The combination of rigid fixation and highly vascular scalp skin reduces the risk of percutaneous driveline infection and may solve an important outstanding problem in use of left ventricular assist devices.
driveline感染限制了左心室辅助装置患者的无事件生存期。随着桥接移植或恢复环境中左心室辅助装置改进前景的不断发展,我们试图降低driveline并发症的风险。
作为牛津Jarvik 2000研究项目的一部分,我们开发了一种碳质然后是钛质的基座,用于将电线穿过皮肤。在绵羊模型中,基座通过肩部皮肤(n = 10)或头皮(n = 9)引出,并在颅骨下方进行固定。仔细检查出口伤口的愈合和感染情况。在另外6只未植入泵的动物中进行了电源线耐久性测试。
累计观察期为1491天(平均时间,78天;范围,14天至198天)。两组之间的观察期没有差异。肩部活动组织中发生感染(n = 2)和愈合受损(n = 5)。安装在颅骨上的基座没有感染或愈合问题。电缆不会因颈部反复弯曲而中断(累计观察期,588天)。当绵羊的头部撞击使碳质基座破裂时,将其替换为钛质基座。
刚性固定和血管丰富的头皮皮肤相结合可降低经皮driveline感染的风险,并可能解决左心室辅助装置使用中的一个重要突出问题。