Min B M, Kim S Y, Yang J M
Department of Ophthalmology, Chungnam National University, Taejon, Korea.
Ophthalmic Surg Lasers. 1998 Feb;29(2):162-6.
Most of the ocular complications of strabismus surgery are related to needle trauma during the suturing of muscles. The authors have developed a muscle-clamping system with an absorbable jaw clip to facilitate exact techniques of eye muscle surgery and to reduce the risk of complications. The absorbable jaw clip is composed of a lower fixing body with three jaws and an upper supporting body. The clip is used to tie an extraocular muscle (which is moved and reattached with the membrane) with a single scleral bite. Using the absorbable jaw clip, the authors performed superior rectus recession on 16 eyes of eight rabbits. The authors examined the conjunctival injection, muscle adhesion power, and light microscopic findings at 1, 2, 4, and 8 weeks postoperatively. The conjunctival injection diminished with time. Bond strengths ranged from 380 to 620 gram gravity, which is sufficient to withstand the normal pull of human extraocular muscles. Some inflammatory cells were found microscopically, and fibrosis increased with time. The muscle-clamping method with the newly invented clip was technically easy, fast, and exact, and it may reduce suturing and complications during strabismus surgery.
斜视手术的大多数眼部并发症与肌肉缝合过程中的针刺创伤有关。作者开发了一种带有可吸收夹钳的肌肉夹闭系统,以促进眼肌手术的精确技术并降低并发症风险。可吸收夹钳由带有三个钳口的下部固定体和上部支撑体组成。该夹钳用于通过单次巩膜咬合来固定一条眼外肌(该肌肉与膜一起移动并重新附着)。作者使用可吸收夹钳对8只兔子的16只眼睛进行了上直肌后徙术。作者在术后1、2、4和8周检查了结膜充血、肌肉附着力和光镜检查结果。结膜充血随时间减轻。粘结强度范围为380至620克力,足以承受人类眼外肌的正常拉力。显微镜下发现了一些炎性细胞,纤维化随时间增加。使用新发明的夹钳进行肌肉夹闭的方法在技术上简便、快速且精确,并且可能减少斜视手术期间的缝合和并发症。