Mario N, Baudin B, Giboudeau J
Service de Biochimie A, Hôpital Saint-Antoine, Paris, France.
J Chromatogr B Biomed Sci Appl. 1998 Feb 27;706(1):123-9. doi: 10.1016/s0378-4347(97)00444-1.
We developed two capillary isoelectric focusing (CIEF) assays, in narrow pH gradients, with the aim of routinely separating and quantitating normal and abnormal hemoglobins (Hbs): a one-step CIEF assay where residual electroosmotic flow mobilizes the proteins during focalization, and a two-step CIEF assay where focused Hbs are mobilized by low pressure by maintaining high-voltage. The resolution of 0.10 pH unit obtained with the one-step assay allowed the separation of the Hbs A, F, S and C; but Hb A2, which represents about 2-3% of whole Hb, could not be quantitated. The better resolution of 0.02 pH unit obtained with the two-step assay allowed the separation of some Hb variants of very close isoelectric points. The reproducibility of retention times was satisfactory (C.V.<5%). Moreover, in this configuration quantitation of Hb A2, Hb F and Hb S led to a standard deviation of less than 5%, allowing the diagnosis of thalassemias. The one-step assay could be useful only for the detection of abnormal variants, while the two-step assay could be applied to the routine analysis of Hbs, with quantitation of minor fractions and presumptive identification of variants.
我们开发了两种窄pH梯度的毛细管等电聚焦(CIEF)测定法,旨在常规分离和定量正常和异常血红蛋白(Hb):一种是一步CIEF测定法,在聚焦过程中残余电渗流使蛋白质移动;另一种是两步CIEF测定法,通过维持高压以低压使聚焦的Hb移动。一步测定法获得的0.10 pH单位的分辨率可分离Hb A、F、S和C;但占总Hb约2-3%的Hb A2无法定量。两步测定法获得的0.02 pH单位的更好分辨率可分离一些等电点非常接近的Hb变体。保留时间的重现性令人满意(变异系数<5%)。此外,在这种配置下,Hb A2、Hb F和Hb S的定量导致标准差小于5%,有助于地中海贫血的诊断。一步测定法仅可用于检测异常变体,而两步测定法可应用于Hb的常规分析,包括微量组分的定量和变体的推定鉴定。