This report describes reprogramming of a pulse generator consequent to cardiac defibrillation. Analysis of the explanted pulse generator revealed normal function. We tested the electrical properties of several defibrillators and our studies suggest that erratic electrostatic discharges at the time of paddle application before defibrillation may create a signal sequence capable of reprogramming the particular pulse generator by activating either the reed switch or internal electronics.

摘要

本报告描述了心脏除颤后脉冲发生器的重新编程情况。对取出的脉冲发生器进行分析显示其功能正常。我们测试了几种除颤器的电性能,我们的研究表明,在除颤前使用电极板时不稳定的静电放电可能会产生一个信号序列,该信号序列能够通过激活簧片开关或内部电子元件来对特定的脉冲发生器进行重新编程。