Validity of the computer science and applications (CSA) activity monitor in children.

作者信息

Trost S G, Ward D S, Moorehead S M, Watson P D, Riner W, Burke J R

机构信息

Department of Exercise Science, University of South Carolina, Columbia, USA.

出版信息

Med Sci Sports Exerc. 1998 Apr;30(4):629-33. doi: 10.1097/00005768-199804000-00023.

Abstract

PURPOSE

The purpose of this study was to evaluate the validity of the CSA activity monitor as a measure of children's physical activity using energy expenditure (EE) as a criterion measure.

METHODS

Thirty subjects aged 10 to 14 performed three 5-min treadmill bouts at 3, 4, and 6 mph, respectively. While on the treadmill, subjects wore CSA (WAM 7164) activity monitors on the right and left hips. VO2 was monitored continuously by an automated system. EE was determined by multiplying the average VO2 by the caloric equivalent of the mean respiratory exchange ratio.

RESULTS

Repeated measures ANOVA indicated that both CSA monitors were sensitive to changes in treadmill speed. Mean activity counts from each CSA unit were not significantly different and the intraclass reliability coefficient for the two CSA units across all speeds was 0.87. Activity counts from both CSA units were strongly correlated with EE (r = 0.86 and 0.87, P < 0.001). An EE prediction equation was developed from 20 randomly selected subjects and cross-validated on the remaining 10. The equation predicted mean EE within 0.01 kcal.min-1. The correlation between actual and predicted values was 0.93 (P < 0.01) and the SEE was 0.93 kcal.min-1.

CONCLUSION

These data indicate that the CSA monitor is a valid and reliable tool for quantifying treadmill walking and running in children.

摘要

目的

本研究旨在以能量消耗(EE)作为标准测量指标,评估CSA活动监测仪作为儿童身体活动测量工具的有效性。

方法

30名年龄在10至14岁的受试者分别以3英里/小时、4英里/小时和6英里/小时的速度在跑步机上进行了三次5分钟的运动。在跑步机上运动时,受试者在左右臀部佩戴CSA(WAM 7164)活动监测仪。通过自动化系统持续监测VO2。通过将平均VO2乘以平均呼吸交换率的热量当量来确定EE。

结果

重复测量方差分析表明,两个CSA监测仪均对跑步机速度的变化敏感。每个CSA单元的平均活动计数无显著差异,两个CSA单元在所有速度下的组内可靠性系数为0.87。两个CSA单元的活动计数均与EE高度相关(r = 0.86和0.87,P < 0.001)。从20名随机选择的受试者中建立了一个EE预测方程,并在其余10名受试者上进行了交叉验证。该方程预测的平均EE误差在0.01千卡·分钟-1以内。实际值与预测值之间的相关性为0.93(P < 0.01),标准误为0.93千卡·分钟-1。

结论

这些数据表明,CSA监测仪是一种有效且可靠的工具,可用于量化儿童在跑步机上行走和跑步的情况。