Haymes E M, Byrnes W C
Human Performance Laboratory, University of Colorado, Boulder 80309.
Med Sci Sports Exerc. 1993 Dec;25(12):1365-9.
The purpose of this study was to examine the accuracy of the Caltrac personal activity computer during walking and running. Ten women and 10 men walked at speeds of 2-5 mph and ran at speeds of 4-8 mph on a horizontal treadmill. Two Caltrac monitors were attached over opposite hips: one programed to give caloric expenditure and the other to give Caltrac counts. Oxygen uptake was measured simultaneously. Significant correlations were found during walking between Caltrac estimated and actual energy expenditure (r = 0.91) and between activity counts and net exercise VO2.kg-1 (r = 0.87). However, the Caltrac significantly overestimated energy cost during horizontal walking at speeds above 2 mph. Although there was a significant correlation between Caltrac estimated and actual energy expenditure during running (r = 0.71), the correlation between Caltrac counts and net exercise VO2.kg-1 was not significant (r = 0.29). There was no significant increase in Caltrac kcal or counts with increased running speed between 5 and 8 mph. It is concluded that the Caltrac is a valid indicator of physical activity during walking but does not adequately discriminate between running speeds of 5-8 mph.
本研究的目的是检验Caltrac个人活动计算机在步行和跑步过程中的准确性。10名女性和10名男性在水平跑步机上以2至5英里/小时的速度行走,以4至8英里/小时的速度跑步。两个Caltrac监测器分别贴在相对的臀部:一个设定为给出热量消耗,另一个给出Caltrac计数。同时测量摄氧量。在步行过程中,发现Caltrac估计的能量消耗与实际能量消耗之间存在显著相关性(r = 0.91),活动计数与净运动VO2.kg-1之间也存在显著相关性(r = 0.87)。然而,Caltrac在速度高于2英里/小时的水平步行过程中显著高估了能量消耗。虽然在跑步过程中Caltrac估计的能量消耗与实际能量消耗之间存在显著相关性(r = 0.71),但Caltrac计数与净运动VO2.kg-1之间的相关性不显著(r = 0.29)。在5至8英里/小时的跑步速度范围内,Caltrac千卡数或计数没有随着跑步速度的增加而显著增加。结论是,Caltrac是步行过程中体力活动的有效指标,但不能充分区分5至8英里/小时的跑步速度。