Cheng T J, Chang H C, Lin T M
Institute of Biomedical Engineering, National Cheng Kung University, Tainan, Taiwan, Republic of China.
Biosens Bioelectron. 1998 Feb 1;13(2):147-56. doi: 10.1016/s0956-5663(97)00111-5.
A Peizoelectric Quartz Crystal (PQC) system possessing an adequate sensitivity to a slight viscosity change is employed to monitor the blood coagulation. In the case of an anticoagulated plasma specimen, prothrombin time (PT) was determined by the PQC sensor on the basis of the introduction of tissue thromboplastin (TF) and calcium ion. Discussing the PQC data statistically, we found that at 90% of the total frequency shift (delta f) the best linear relationship in a double-logarithmic plot of PT versus TF concentration in the range of 5.466-22.311 unit/ml is shown. The PQC result was compared with commercial optical coagulometry and showed a good linear correlation (ca. 0.98). The clotting and recalcification of whole blood were also studied by the PQC system. The testing time for one measurement could be completed within 20 min. and had a good agreement with that obtained conventionally.
采用对轻微粘度变化具有足够灵敏度的压电石英晶体(PQC)系统来监测血液凝固。对于抗凝血浆样本,通过在引入组织凝血活酶(TF)和钙离子的基础上,利用PQC传感器测定凝血酶原时间(PT)。对PQC数据进行统计学分析时,我们发现在总频率偏移(δf)的90%时,PT与TF浓度在5.466 - 22.311单位/毫升范围内的双对数图中呈现出最佳线性关系。将PQC结果与商用光学凝固法进行比较,显示出良好的线性相关性(约为0.98)。还通过PQC系统研究了全血的凝血和再钙化过程。一次测量的测试时间可在20分钟内完成,并且与传统方法获得的结果具有良好的一致性。