Suppr超能文献

用于毛细管电泳的微型热透镜装置。

Miniaturized thermal lens device for capillary electrophoresis.

作者信息

Seidel B S, Faubel W

机构信息

Forschungszentrum Karlsruhe, Institut für Instrumentelle Analytik, Germany.

出版信息

Biomed Chromatogr. 1998 May-Jun;12(3):155-7. doi: 10.1002/(SICI)1099-0801(199805/06)12:3<155::AID-BMC793>3.0.CO;2-R.

Abstract

In this paper we will report on the combination of a miniaturized fiber optic modified photothermal sensor (size 2 x 2 x 2 cm) to capillary electrophoresis (CE) as detection device. The properties and the performances of the thermal lens detector, based on a double beam absorption scheme, were studied in a capillary electrophoresis system with different types of toxic pollutants, e.g. pesticides. A tunable Argon ion laser, which emits between 364 and 514 nm, was adapted to a commercial injection and high voltage CE system. The miniaturization of the thermal lens detector head with optical waveguides allows the combination with typical CE systems, therefore pesticides in microliter volumes can be treated. The detector performance is demonstrated with the monitoring of pesticides in capillaries with various diameter. The limit of detection is one of two orders of magnitude better than conventional absorption detector.

摘要

在本文中,我们将报道一种小型化光纤改性光热传感器(尺寸为2×2×2厘米)与作为检测装置的毛细管电泳(CE)的结合。基于双光束吸收方案的热透镜检测器的特性和性能,在具有不同类型有毒污染物(如农药)的毛细管电泳系统中进行了研究。一台发射波长在364至514纳米之间的可调谐氩离子激光器,被适配到一个商用进样和高压CE系统中。热透镜检测头与光波导的小型化使得其能够与典型的CE系统相结合,因此可以处理微升体积的农药。通过监测不同直径毛细管中的农药,展示了该检测器的性能。其检测限比传统吸收检测器好两个数量级。