Suppr超能文献

Comparison of percent total GHb with percent HbA1c in people with and without known diabetes.

作者信息

Nuttall F Q

机构信息

Minneapolis Veterans Affairs Medical Center, Department of Medicine, University of Minnesota, 55417, USA.

出版信息

Diabetes Care. 1998 Sep;21(9):1475-80. doi: 10.2337/diacare.21.9.1475.

Abstract

OBJECTIVE

To directly compare results obtained using an ion-exchange high-performance liquid chromatography (HPLC) HbA1c method used in the Diabetes Control and Complications Trial with two different affinity chromatography methods in which "total GHb" is determined.

RESEARCH DESIGN AND METHODS

Blood was obtained from a large number of people with and without known diabetes. The specimens were divided and assayed for HbA1c and for total GHb. Total GHb was determined using a semi-automated gravity-elution boronate affinity chromatography method and an automated boronate affinity HPLC method. The results obtained with the two methods were also compared.

RESULTS

In subjects without known diabetes, the mean percentage HbA1c and the range of values were similar to the total GHb values in the same subjects when assayed using the semi-automated affinity gravity-elution method (mean 5.2 +/- 0.4 and 5.1 +/- 0.4% [SD], respectively). With the affinity HPLC method, results were 5.3 +/- 0.4%. The similarity in results was surprising. However, analysis of the data suggests that a large proportion of the material in the HbA1c fraction measured using this ion-exchange HPLC method is not GHb, as pointed out by others. Although the results were similar in people without known diabetes, in the people with diabetes, the incremental increase was approximately 25% greater for the total GHb when compared with the increase in HbA1c. When corrected for the non-GHb being measured by the HbA1c method, it can be calculated that approximately 40% more GHb is measured using affinity chromatography over the entire range of GHb values.

CONCLUSIONS

The similarity in the mean and range of percent HbA1c and in percent total GHb using these different methods can be attributed to two factors: 1) the HbA1c ion-exchange method measures only approximately 50-60% of the total GHb present, and 2) approximately 40-50% of the material being measured in the HbA1c fraction is not GHb, i.e., offsetting factors fortuitously resulted in values similar to the more specific affinity methods. The greater incremental increase in percent total GHb compared with percent HbA1c in people with diabetes can be attributed to the greater amount of GHb being measured with the affinity methods.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验