Dave T, Ruston D, Fowler C
Department of Vision Sciences, University of Aston, Birmingham, United Kingdom.
Optom Vis Sci. 1998 Sep;75(9):647-55. doi: 10.1097/00006324-199809000-00019.
To compare the agreement and repeatability of keratometry measurements with simulated keratometry measurements from the EyeSys (model II) corneal analysis system (CAS). Furthermore, to establish any relationship between the repeatability of peripheral corneal measurements and the spatial location at the cornea.
Measurements using a Bausch & Lomb keratometer in 19 eyes (19 subjects) were compared with the simulated keratometry values using the EyeSys CAS. Repeatability of the keratometer (14 eyes, 14 subjects) and the EyeSys simulated keratometry values (10 eyes, 10 subjects) was assessed by comparing the difference of two measurements. Peripheral repeatability of the EyeSys CAS was compared in a similar manner in a sample of 10 eyes (10 subjects).
The 95% confidence limits (-0.529 to +0.149 D) showed a lack of agreement between both instruments. With respect to the repeatability, both the keratometer (SD = +/- 0.103 D) and the EyeSys (SD = +/- 0.072 D) were found to exhibit approximately similar degrees of repeatability. The repeatability of peripheral corneal measurements using the EyeSys CAS showed changes in repeatability dependent on the corneal meridian and the distance from the point of alignment. Measurements along the superior and nasal meridians showed poorest repeatability. Repeatability was also found to deteriorate away from the point of alignment.
EyeSys simulated keratometry values were not interchangeable with the keratometer. This could be due to differences in the method of alignment or measurements from different areas on the corneal surface. Repeatability of both the Bausch & Lomb keratometer and the EyeSys CAS was found to be similar. The spatial dependency of peripheral corneal radius repeatability measurements could be attributed to interference of the ocular adnexa in those areas of measurement.
比较使用Bausch & Lomb角膜曲率计测量的角膜曲率与EyeSys(II型)角膜分析系统(CAS)模拟的角膜曲率测量值之间的一致性和可重复性。此外,确定周边角膜测量的可重复性与角膜上空间位置之间的任何关系。
将19只眼(19名受试者)使用Bausch & Lomb角膜曲率计的测量结果与使用EyeSys CAS模拟的角膜曲率值进行比较。通过比较两次测量的差异,评估角膜曲率计(14只眼,14名受试者)和EyeSys模拟角膜曲率值(10只眼,10名受试者)的可重复性。以类似方式在10只眼(10名受试者)的样本中比较EyeSys CAS的周边可重复性。
95%置信限(-0.529至+0.149 D)表明两种仪器之间缺乏一致性。关于可重复性,发现角膜曲率计(标准差=±0.103 D)和EyeSys(标准差=±0.072 D)表现出大致相似的可重复性程度。使用EyeSys CAS进行周边角膜测量的可重复性显示,可重复性随角膜子午线和距对准点的距离而变化。沿上方和鼻侧子午线的测量显示出最差的可重复性。还发现可重复性在远离对准点时变差。
EyeSys模拟的角膜曲率值与角膜曲率计不可互换。这可能是由于对准方法或角膜表面不同区域测量的差异。发现Bausch & Lomb角膜曲率计和EyeSys CAS的可重复性相似。周边角膜半径可重复性测量的空间依赖性可归因于眼附属器在那些测量区域的干扰。