Mehravaran Shiva, Asgari Soheila, Bigdeli Sara, Shahnazi Ashkan, Hashemi Hassan
Noor Ophthalmology Research Center, Noor Eye Hospital, 96 Esfandiar Blvd., Vali'asr Ave., Tehran, Iran.
Int Ophthalmol. 2014 Aug;34(4):869-75. doi: 10.1007/s10792-013-9895-3. Epub 2014 Feb 23.
The purpose of this study was to determine and compare the repeatability of keratometry measurements with the Javal manual keratometer, Topcon automated kerato-refractometer, IOLMaster, EyeSys Corneal Analysis System, and Pentacam Scheimpflug Topography System, and determine the agreement of measurements of the latter four devices with the Javal keratometer as the gold standard. In this cross-sectional study, 21 people with no history of ocular disease or surgery were examined twice with each device. Minimum, maximum, and average keratometry readings were recorded in diopters (D) for each eye and used in the analyses. For statistical analysis, we determined correlation coefficients and used the Bland-Altman method and calculated the 95 % limits of agreement (LoA). All repeatability coefficients were satisfactorily high. Best repeatability for minimum keratometry and maximum keratometry readings was seen with IOLMaster (95 % LoA -0.23 to 0.19 D and -0.31 to 0.32 D, respectively). Agreement with Javal manual keratometry was best with Topcon and IOLMaster for minimum keratometry readings (95 % LoA -0.67 to 0.28 D and -0.57 to 0.38 D, respectively), and with Topcon for maximum keratometry readings (95 % LoA -0.85 to 0.63 D). In our series of normal eyes, keratometry readings with Topcon, IOLMaster, and Pentacam showed very good agreement with Javal measurements, and inter-device agreements were better than the repeatability of the manual gold standard. However, differences between Javal and EyeSys may not be clinically acceptable, and these two devices should not be used interchangeably. Further studies are needed to investigate these issues in non-virgin eyes.
本研究的目的是确定并比较使用Javal手动角膜曲率计、拓普康自动角膜屈光计、IOLMaster、EyeSys角膜分析系统和Pentacam Scheimpflug角膜地形图系统进行角膜曲率测量的可重复性,并将后四种设备的测量结果与作为金标准的Javal角膜曲率计的测量结果进行一致性分析。在这项横断面研究中,对21名无眼部疾病或手术史的受试者使用每种设备进行了两次检查。记录每只眼睛以屈光度(D)为单位的最小、最大和平均角膜曲率读数,并用于分析。对于统计分析,我们确定了相关系数,并使用Bland-Altman方法计算了95%一致性界限(LoA)。所有可重复性系数都令人满意地高。IOLMaster在最小角膜曲率和最大角膜曲率读数方面的可重复性最佳(95% LoA分别为-0.23至0.19 D和-0.31至0.32 D)。在最小角膜曲率读数方面,拓普康和IOLMaster与Javal手动角膜曲率测量的一致性最佳(95% LoA分别为-0.67至0.28 D和-0.57至0.38 D),在最大角膜曲率读数方面,拓普康与Javal手动角膜曲率测量的一致性最佳(95% LoA为-0.85至0.63 D)。在我们的正常眼系列中,拓普康、IOLMaster和Pentacam的角膜曲率读数与Javal测量结果显示出非常好的一致性,并且设备间的一致性优于手动金标准的可重复性。然而,Javal和EyeSys之间的差异可能在临床上不可接受,这两种设备不应互换使用。需要进一步的研究来调查非初次检查眼睛中的这些问题。