Poller L, van den Besselaar A M, Jespersen J, Tripodi A, Houghton D
Central Facility, ECAA, The University of Manchester, UK.
Thromb Haemost. 1998 Oct;80(4):615-23.
Local calibration studies have been performed with lyophilized plasmas at 155 European laboratories to assess coagulometer effects on manual ISI of ECAA thromboplastins (low ISI human recombinant and high ISI rabbit brain). Common sets of 7 normal and 20 artificially depleted lyophilized plasmas were tested with the thromboplastins in the routine local coagulometers. With the human reagent, marked lowering of the manual ISI resulted with most coagulometers, which was associated with disproportionate shortening of the normals. Where this shortening did not occur, there was little coagulometer effect on the ISI. With the rabbit reagent, proportionate shortening with both normal and abnormal plasmas occurred in most instruments with little effect on ISI. INR correction by local ISI assignment appeared successful with the ECAA human reagent. There was negligible INR correction from local calibration of the ECAA rabbit reagent in coagulometers where thromboplastin ISI were unchanged from the ECAA established manual values.
在155家欧洲实验室,已使用冻干血浆进行了局部校准研究,以评估凝血仪对欧洲临床化学协会(ECAA)凝血活酶(低国际敏感度指数(ISI)的人重组凝血活酶和高ISI的兔脑凝血活酶)手工ISI的影响。使用这些凝血活酶在常规局部凝血仪中对7份正常和20份人工耗尽的冻干血浆的通用组进行了检测。对于人源试剂,大多数凝血仪导致手工ISI显著降低,这与正常样本不成比例地缩短有关。在未出现这种缩短的情况下,凝血仪对ISI几乎没有影响。对于兔源试剂,大多数仪器中正常和异常血浆均出现成比例缩短,对ISI影响很小。通过局部ISI赋值进行国际标准化比值(INR)校正,对于ECAA人源试剂似乎是成功的。在凝血活酶ISI与ECAA既定手工值无变化的凝血仪中,通过ECAA兔源试剂的局部校准进行INR校正可忽略不计。