Sutton J G, Bosley C, Rands A
Forensic Science Service, London Laboratory, United Kingdom.
Sci Justice. 1998 Jul-Sep;38(3):157-64. doi: 10.1016/S1355-0306(98)72100-8.
The potential of specific human semen markers, P30 and 19-OH prostaglandin F1 alpha/F2 alpha (PG F), ELISAs were compared in a blind trial. The PG F assay was shown to be at least ten times more sensitive than the P30 technique although cross reactivity problems with spermicidal creams and other possible domestic contaminants were identified. A simple method for the removal of such contaminants using a phenyl boronate column (PBA) was evaluated in an extensive blind trial.