Validation of an automated microplate enzyme immunoassay for screening of postmortem blood for drugs of abuse.

作者信息

Spiehler V R, Collison I B, Sedgwick P R, Perez S L, Le S D, Farnin D A

机构信息

Spiehler & Associates, Newport Beach, California, USA.

出版信息

J Anal Toxicol. 1998 Nov-Dec;22(7):573-9. doi: 10.1093/jat/22.7.573.

Abstract

The objective of this study was to compare the sensitivity and specificity of an enzyme immunoassay employing antibodies bound to a microtiter plate (MPEIA) with those of two radioimmunoassays for screening postmortem blood from selected coroner's cases for drugs of abuse. The radioimmunoassays were a coated-tube radioimmunoassay (CTRIA) and a double antibody radioimmunoassay (DARIA). Specimens consisted of 260 postmortem blood specimens from coroner's cases. Immunoassay results (positive or negative) were compared with confirmed results on those cases by gas chromatography-mass spectrometry, alone or in combination with gas-liquid chromatography using either a nitrogen-phosphorus or flame-ionization detector. Sensitivity was calculated as the true-positive rate using chromatographic confirmation as the reference standard. Specificity was calculated as the true-negative rate. Sensitivity and specificity were calculated for 5-7 potential cutoff concentrations for the drug classes opiates, amphetamines, cocaine and metabolites, and barbiturates. For opiates, the sensitivity and specificity were 99% and 93%, respectively, for the MPEIA at a cutoff of 20-ng/mL morphine, compared with 94% and 96% for the CTRIA at a cutoff of 5-ng/mL morphine and > 99% and 96% for the DARIA at 20-ng/mL morphine. For cocaine and metabolites, the sensitivity and specificity were 96% and 93%, respectively, for the MPEIA at 50-ng/mL benzoylecgonine, compared with 93% and 96% for CTRIA at 50-ng/mL benzoylecgonine and 98% and 97% for the DARIA at 50-ng/mL benzoylecgonine. For amphetamines, the sensitivity and specificity were >99% and 91%, respectively, for the MPEIA at 25-ng/mL methamphetamine, compared with 93% and 86% for the CTRIA at 25-ng/mL methamphetamine and 83% and 89% for the DARIA at 50-ng/mL methamphetamine. For barbiturates, the sensitivity and specificity were > 99% and 92%, respectively, for the MPEIA at 50-ng/mL secobarbital, compared with 91% and 87% for the CTRIA at 500-ng/mL secobarbital and 79% and 95% for the DARIA at a cutoff of 1000-ng/mL phenobarbital.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验