Comparison of methods to assess visual impairment from glare and light scattering with posterior capsule opacification.

作者信息

Tan J C, Spalton D J, Arden G B

机构信息

Department of Ophthalmology, St. Thomas' Hospital, London, United Kingdom.

出版信息

J Cataract Refract Surg. 1998 Dec;24(12):1626-31. doi: 10.1016/s0886-3350(98)80354-4.

Abstract

PURPOSE

To compare 2 glare tests to determine their relative usefulness in the assessment of posterior capsule opacification (PCO) and to evaluate the potential benefits of combined visual, acuity, contrast sensitivity, and glare testing.

SETTING

Teaching hospital ophthalmology department.

METHODS

Sixteen patients had glare, visual acuity, and contrast sensitivity testing before and after neodymium:YAG (Nd:YAG) capsulotomy. Results with the Brightness Acuity Tester (BAT, Mentor), which measures disability glare, and the Straylightmeter (Foundation for Eye Research, The Netherlands), which quantifies forward scatter by direct compensation techniques, were compared. The correlation between glare, ETDRS visual acuity, and Pelli-Robson contrast sensitivity was determined.

RESULTS

Pretreatment visual acuity was significantly correlated with contrast sensitivity (P < .01). However, visual acuity and contrast sensitivity were poorly correlated with both the BAT and Straylightmeter (P > .05), indicating that visual acuity is predictive of contrast sensitivity but a poor predictor of glare. Glare was significantly improved (Straylightmeter, P < .0001; BAT, P < .05) following capsulotomy. While the Straylightmeter consistently measured precapsulotomy forward scatter that improved with treatment, corresponding BAT disability glare was unmeasurable in 18.8% of patients with PCO, as their visual acuities improved rather than deteriorated with glare testing.

CONCLUSIONS

Glare testing provided more information than contrast sensitivity when combined with visual acuity in the evaluation of PCO. Glare related to PCO is better assessed using the Straylightmeter because the BAT may yield aberrant disability glare results.

摘要

目的

比较两种眩光测试,以确定它们在评估后囊膜混浊(PCO)方面的相对效用,并评估综合视力、视敏度、对比敏感度和眩光测试的潜在益处。

设置

教学医院眼科。

方法

16例患者在钕:钇铝石榴石(Nd:YAG)囊切开术前和术后进行了眩光、视力和对比敏感度测试。比较了测量失能眩光的亮度视力测试仪(BAT,美敦力公司)和通过直接补偿技术量化前向散射的散射光计(荷兰眼科研究基金会)的结果。确定了眩光、ETDRS视力和佩利-罗布森对比敏感度之间的相关性。

结果

治疗前视力与对比敏感度显著相关(P <.01)。然而,视力和对比敏感度与BAT和散射光计的相关性均较差(P >.05),表明视力可预测对比敏感度,但对眩光的预测性较差。囊切开术后眩光显著改善(散射光计,P <.0001;BAT,P <.05)。虽然散射光计始终测量到囊切开术前的前向散射随治疗而改善,但在18.8%的PCO患者中,相应的BAT失能眩光无法测量,因为他们的视力在眩光测试中有所提高而非恶化。

结论

在评估PCO时,眩光测试与视力相结合比对比敏感度提供了更多信息。使用散射光计能更好地评估与PCO相关的眩光,因为BAT可能会产生异常的失能眩光结果。