Schürmann K, Vorwerk D, Bücker A, Grosskortenhaus S, Günther R W
Department of Diagnostic Radiology, University of Technology, Aachen, Germany.
Cardiovasc Intervent Radiol. 1998 Sep-Oct;21(5):411-8. doi: 10.1007/s002709900289.
To compare patency and neointima formation of single and tandem arterial stents.
In each of six sheep, two Memotherm nitinol stents (tandem stents) were inserted into the external iliac artery on one side and a single stent into the artery on the opposite side. The size of the iliac lumen was assessed in the proximal, middle, and distal segments of the stents by intravascular ultrasound (IVUS) before, immediately after, and 1 month after implantation when the sheep were killed. Neointimal thickness was determined in the proximal, middle, and distal segments of each stent by light microscopy.
All stents remained patent. There was no significant difference in lumen and neointimal thickness between single and tandem stents. Cranial tandem stents showed a significantly wider lumen and smaller neointimal thickness than caudal tandem stents. In the proximal and distal segments, the lumen of the stents was significantly smaller and the neointimal thickness greater than in the middle segment; differences in neointimal thickness were significant only between the proximal and the middle segment.
In an experimental setting, tandem stents did not interfere with one another with regard to patency and neointima formation when compared with a single contralateral stent. Neointimal thickening after stent insertion seems to be inversely related to the original arterial diameter.
比较单个动脉支架和串联动脉支架的通畅性及新生内膜形成情况。
在6只绵羊中,于一侧髂外动脉植入两个Memotherm镍钛合金支架(串联支架),另一侧髂外动脉植入单个支架。在植入前、植入后即刻以及绵羊处死时(植入后1个月),通过血管内超声(IVUS)评估支架近端、中段和远端的髂管腔大小。通过光学显微镜测定每个支架近端、中段和远端的新生内膜厚度。
所有支架均保持通畅。单个支架和串联支架在管腔和新生内膜厚度方面无显著差异。头端串联支架显示出比尾端串联支架更宽的管腔和更小的新生内膜厚度。在近端和远端节段,支架管腔显著小于中段,新生内膜厚度大于中段;仅近端和中段之间新生内膜厚度差异显著。
在实验环境中,与对侧单个支架相比,串联支架在通畅性和新生内膜形成方面互不干扰。支架植入后新生内膜增厚似乎与原始动脉直径呈负相关。