Pena A, Carmona A, Barbosa A, Lino C, Silveira I, Castillo B
Laboratory of Bromatology, Faculty of Pharmacy, University of Coimbra, Portugal.
J Pharm Biomed Anal. 1998 Dec;18(4-5):839-45. doi: 10.1016/s0731-7085(98)00268-4.
A liquid chromatographic method of tetracycline and its major degradation products on a C8-reversed phase column with acidic mobile phase and fluorescence detection is described. The quantification limit, measured as the amount of sample that gave a signal ten times the peak-to-peak noise of the baseline, was: 0.25 ng for tetracycline (TC) and epitetracycline (ETC), 25 ng for and 4-epianhydrotetracycline (EATC) and 50 ng for anhydrotetracycline (ATC) of injected standard. By means of this liquid chromatography (LC) assay TC, ETC, EATC and ATC as main degradation products of tetracycline, can be separated and determined with good sensitivity and specificity within 15 min.
描述了一种在C8反相柱上采用酸性流动相和荧光检测法测定四环素及其主要降解产物的液相色谱方法。以产生的信号为基线峰-峰噪声10倍的样品量计算,定量限为:注入标准品时,四环素(TC)和差向四环素(ETC)为0.25 ng,4-表脱水四环素(EATC)为25 ng,脱水四环素(ATC)为50 ng。通过这种液相色谱(LC)测定法,作为四环素主要降解产物的TC、ETC、EATC和ATC能够在15分钟内以良好的灵敏度和特异性进行分离和测定。