Image analysis of implanted rigid and foldable intraocular lenses in human eyes using Scheimpflug photography.

作者信息

Klos K M, Richter R, Schnaudigel O, Ohrloff C

机构信息

University Eye Hospital, Johann Wolfgang Goethe University, Frankfurt, Germany.

出版信息

Ophthalmic Res. 1999;31(2):130-3. doi: 10.1159/000055524.

Abstract

We evaluated the quality of imaging posterior chamber intraocular lenses (PC IOL) by Scheimpflug photography. We took photographs of 190 rigid or foldable PC IOLs using the Scheimpflug system EAS-1000 from Nidek and provided digital image analysis. We found a good imaging of PC IOLs and the cristalline lens capsule, data obtained by digital image analysis were reliable and reproducible and Scheimpflug photography could be demonstrated as a sensitive tool to show any irregularities of IOL material. Thus, we found in foldable IOLs 'glistenings', which did not have any influence on functional results and we could distinguish them from lens damage. We propose to use Scheimpflug tests regularly for the examination of implanted new types of IOLs.

摘要

我们通过Scheimpflug摄影术评估了后房型人工晶状体(PC IOL)的成像质量。我们使用Nidek公司的EAS - 1000 Scheimpflug系统对190枚硬性或可折叠式PC IOL进行拍照,并进行数字图像分析。我们发现PC IOL和晶状体囊膜成像良好,数字图像分析所获得的数据可靠且可重复,Scheimpflug摄影术可作为一种灵敏的工具来显示IOL材料的任何不规则情况。因此,我们在可折叠式IOL中发现了“闪烁现象”,其对功能结果没有任何影响,并且我们能够将它们与晶状体损伤区分开来。我们建议定期使用Scheimpflug测试来检查植入的新型IOL。