Suppr超能文献

一种新型pH/血气质量控制系统与其他两种方法相比的长期可重复性。

Long-term reproducibility of a new pH/blood-gas quality-control system compared to two other procedures.

作者信息

Noonan D C, Komjathy Z L

出版信息

Clin Chem. 1976 Nov;22(11):1817-20.

PMID:10098
Abstract

The long-term precision and stability of a new quality-control system for blood pH and gas measurements are compared to that of tonometered bicarbonate solutions and serum-based preparations. The new system, consisting of gas-equilibrated bicarbonate solutions in glass ampuls, is shown to be as stable as the serum-based preparation, and as reproducible as either of the other methods. The new system, offering three discrete sets of control values, has certain advantages in the simultaneous quality of pH, carbon dioxide tension, and oxygen tension measurements.

摘要

将一种用于血液pH值和气体测量的新质量控制系统的长期精度和稳定性与用血气分析仪校准的碳酸氢盐溶液和基于血清的制剂进行了比较。新系统由玻璃安瓿中的气体平衡碳酸氢盐溶液组成,结果表明其稳定性与基于血清的制剂相当,重现性与其他两种方法中的任何一种相当。新系统提供三组离散的控制值,在同时测量pH值、二氧化碳分压和氧分压方面具有一定优势。