Stability and precision of a new ampuled quality-control system for pH and blood-gas measurements.

作者信息

Komjathy Z L, Mathies J C, Parker J A, Schreiber H A

出版信息

Clin Chem. 1976 Aug;22(8):1399-401.

PMID:7369
Abstract

We describe an evaluation of the in-use stability and short-term precision of a three-level ampuled quality-control system for monitoring pH, pCO2, and pO2 measurements on clinical blood-gas analyzers. In three hospital laboratories, 324 such ampuls were opened and allowed to stand with their contents exposed to atmospheric conditions for accurately timed intervals up to 240 s. Contents were then analyzed for pH, pCO2, and pO2. Student's t-test was used to evaluate the significance of differences observed in recoveries after time exposure. At a signifcance level of P less than or equal to 0.05, the only significant changes observed throughout the first minute of exposure were average pO2 increases of 180 Pa (1.4 mmHg) (+ 1.4%) and 230 Pa (1.7 mmHg) (+ 2.9%) at levels of 13.4 and 7.7 kPa kPa (101 and 58 mmHg), respectively. The ampuled system was found to be stable precise convenient, and suitable for use in the routine laboratory.

摘要

我们描述了一种用于监测临床血气分析仪上pH、pCO2和pO2测量值的三级安瓿瓶质量控制系统的使用稳定性和短期精密度评估。在三家医院实验室中,打开了324个此类安瓿瓶,并让其内容物在大气条件下放置精确的时间间隔,最长可达240秒。然后分析内容物的pH、pCO2和pO2。采用学生t检验来评估暴露时间后回收率差异的显著性。在显著性水平P小于或等于0.05时,在暴露的第一分钟内观察到的唯一显著变化是,在13.4和7.7 kPa(101和58 mmHg)水平下,平均pO2分别增加了180 Pa(1.4 mmHg)(+1.4%)和230 Pa(1.7 mmHg)(+2.9%)。发现该安瓿瓶系统稳定、精确、方便,适用于常规实验室。