Electronic length measurement using small and large files in enlarged canals.

作者信息

Nguyen H Q, Kaufman A Y, Komorowski R C, Friedman S

机构信息

Department of Endodontics, Faculty of Dentistry, University of Toronto, Ontario, Canada.

出版信息

Int Endod J. 1996 Nov;29(6):359-64. doi: 10.1111/j.1365-2591.1996.tb01398.x.

Abstract

Electronic apex locators are frequently used attached to a small size endodontic file; however, the effect on the measurement of the relative diameters of the file and the root canal has not been clarified. In this study, the length of enlarged canals was measured using small size files and files matching the canal diameter to observe a possible discrepancy. The accuracy of electronic length control during canal preparation with rotary files was also assessed. The root canals in 21 extracted, single rooted teeth were accessed, and their actual length (AL) established by passing a size 10 file just through the minor apical foramen. The teeth were then embedded in an alginate mold. The initial canal length (IL) was measured with the Root ZX apex locator by negotiating a size 10 file to the apical constriction. The canal was enlarged to size 60 with rotary files while the length was continuously controlled with the apex locator. Then, the final length measurements were obtained with a size 10 file and a size 60 file (FL-10 and FL-60, respectively). The average values of IL, FL-10 and FL-60 were calculated and compared using Repeated Measures Analysis of Variance followed by Tukey's Studentized Range test (P < 0.05). Using composite resin, the size 60 files were secured at the FL-60 length, the teeth were removed from the alginate mold, stained with Picroformal DI Buoin stain and the position of the file tip was observed histomorphometrically after the apical 4 mm of the canal was exposed by grinding the buccal aspect of the root. The IL was 0.45 mm shorter than AL (P < 0.05). The differences between FL-10, FL-60 and IL were not statistically significant. Histomorphometrically, the apical constriction was absent in all the teeth, but the file tips were confined within the root. This study concluded that the Root ZX indicated the location of an apical constriction even when the anatomic constriction was eliminated. In the enlarged canals, length measurements obtained with small and large size files were comparable.

摘要

电子根尖定位仪常与小尺寸根管锉配套使用;然而,其对根管锉和根管相对直径测量的影响尚未明确。在本研究中,使用小尺寸锉和与根管直径匹配的锉测量扩大根管的长度,以观察可能存在的差异。同时还评估了使用旋转锉进行根管预备时电子长度控制的准确性。选取21颗拔除的单根牙,开通根管,通过一根10号锉刚好穿过根尖小孔确定其实际长度(AL)。然后将牙齿嵌入藻酸盐模型中。使用Root ZX根尖定位仪,通过将一根10号锉插入根尖狭窄处测量初始根管长度(IL)。使用旋转锉将根管扩大至60号,同时用根尖定位仪持续控制长度。然后,分别用10号锉和60号锉获得最终长度测量值(分别为FL-10和FL-60)。计算IL、FL-10和FL-60的平均值,并使用重复测量方差分析和Tukey's Studentized Range检验进行比较(P<0.05)。使用复合树脂将60号锉固定在FL-60长度处,将牙齿从藻酸盐模型中取出,用苦味酸-甲醛-DI Buoin染色剂染色,在磨除牙根颊侧使根管根尖4mm暴露后,通过组织形态计量学观察锉尖的位置。IL比AL短0.45mm(P<0.05)。FL-10、FL-60与IL之间的差异无统计学意义。组织形态计量学显示,所有牙齿均无根尖狭窄,但锉尖均位于牙根内。本研究得出结论,即使解剖学上的狭窄消失,Root ZX仍能指示根尖狭窄的位置。在扩大的根管中,使用小尺寸和大尺寸锉获得的长度测量值具有可比性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验