Non-invasive temperature imaging of muscles with magnetic resonance imaging using spin-echo sequences.

作者信息

Mietzsch E, Koch M, Schaldach M, Werner J, Bellenberg B, Wentz K U

机构信息

Department of Biomedical Engineering, Medical Faculty, Ruhr-University, Bochum, Germany.

出版信息

Med Biol Eng Comput. 1998 Nov;36(6):673-8. doi: 10.1007/BF02518868.

Abstract

The application of spin-echo magnetic resonance imaging sequences on non-invasive temperature imaging for temperature mapping of human limbs is investigated. In an in vitro experiment performed on a meat sample, the equilibrium magnetisation P and the spin-lattice relaxation time T1 are calculated from the values for the repetition time TR and the signal intensities obtained by a spin-echo sequence at different tissue temperatures as measured by a fibre-optic probe. T1 is linearly correlated to the tissue temperature, and P is linearly correlated to the reciprocal value of the absolute temperature. Both effects, taken together, lead to a non-linear dependency of the signal intensity on temperature. Therefore a TR leading to maximum temperature dependency of the signal intensity is calculated and used in the further experiments. In the in vivo experiments, the lower legs of two volunteers are cooled from outside. Images are acquired with a spin-echo sequence (1.5 T, TR = 1200 ms, TE = 10 ms). A rise in signal intensity in the muscle with falling skin temperature is observed, particularly in more peripheral muscle layers. This study shows that spin-echo sequences can be used to monitor temperature changes and temperature differences in living muscle tissue.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验