Young P J, Blunt M C
Intensive Care Unit, Queen Elizabeth Hospital, Kings Lynn, UK.
Crit Care. 1999;3(5):123-6. doi: 10.1186/cc357. Epub 1999 Aug 31.
The Portex Soft Seal high-volume, low-pressure cuffed tracheal tube was compared with the Mallinckrodt HiLo, Sheridan Preformed and Portex Profile tracheal tubes for leakage of dye placed in the subglottic space of a pig's trachea which was used in a benchtop mechanical ventilation model and in six isolated pig tracheas.
There was no leakage, either in the ventilation model or in the isolated tracheas in the Portex Soft Seal group. There was rapid leakage in the ventilation model and in all the isolated tracheas for the Mallinckrodt HiLo, and five out of six isolated tracheas for the Sheridan Preformed and the Portex Profile group.
This benchtop study suggests that the improved compliance characteristics of the Portex Soft Seal cuff are beneficial in preventing leakage of fluid in these models.
在一个台式机械通气模型和六个离体猪气管中,将波特克斯软密封大容量、低压带套囊气管导管与马林克罗特高-低型、谢里登预制型和波特克斯轮廓型气管导管进行比较,观察置于猪气管声门下间隙的染料泄漏情况。
波特克斯软密封组在通气模型和离体气管中均无泄漏。马林克罗特高-低型在通气模型和所有离体气管中均迅速出现泄漏,谢里登预制型和波特克斯轮廓型组在六个离体气管中有五个出现泄漏。
这项台式研究表明,波特克斯软密封套囊改善的顺应性特征有利于在这些模型中防止液体泄漏。