Stent longitudinal flexibility: a comparison of 13 stent designs before and after balloon expansion.

作者信息

Ormiston J A, Dixon S R, Webster M W, Ruygrok P N, Stewart J T, Minchington I, West T

机构信息

Auckland Heart Group, Mercy Angiography, Auckland, New Zealand.

出版信息

Catheter Cardiovasc Interv. 2000 May;50(1):120-4. doi: 10.1002/(sici)1522-726x(200005)50:1<120::aid-ccd26>3.0.co;2-t.

Abstract

Longitudinal flexibility is an important property of coronary stents, facilitating delivery and allowing the expanded stent to conform to vessel contour. Subjective descriptions of flexibility abound, but there are few independent quantitative data to aid stent selection. A three-point bend test was employed to measure stiffness, the reciprocal of flexibility, for 13 stent designs in the unexpanded (bare) state, then after expansion with a 3.5-mm balloon. For eight of the designs, stiffness of the proprietary stent/balloon delivery system was also measured. In the unexpanded state, there was a wide spread of stiffness, which ranged from 0.5+/-0.2 to 91.5+/-10.0 g force/mm, depending on design. Stiffness was least for the coil (Wiktor and Crossflex) and hybrid (AVE GFX and Bard XT) designs. The MultiLink was the most flexible and the Crown the stiffest of the slotted tube designs. All stents became stiffer upon expansion. For most manufacturer-mounted stents, the delivery balloon was the main determinant of stent/balloon delivery system stiffness. Manufacturer-mounted stent profile ranged from 1.15+/-0.11 mm for the Jostent to 1.53 +/- 0.05 mm for the MultiLink system. Independent quantitative assessment of characteristics such as flexibility and profile should aid rational comparison of stent designs.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验