Prieto P M, Vargas-Martín F, Goelz S, Artal P
Departamento de Física, Universidad de Murcia, Spain.
J Opt Soc Am A Opt Image Sci Vis. 2000 Aug;17(8):1388-98. doi: 10.1364/josaa.17.001388.
A description of a Hartmann-Shack sensor to measure the aberrations of the human eye is presented. We performed an analysis of the accuracy and limitations of the sensor using experimental results and computer simulations. We compared the ocular modulation transfer function obtained from simultaneously recorded double-pass and Hartmann-Shack images. The following factors affecting the sensor performance were evaluated: the statistical accuracy, the number of modes used to reconstruct the wave front, the size of the microlenses, and the exposure time.
本文介绍了一种用于测量人眼像差的哈特曼-夏克传感器。我们利用实验结果和计算机模拟对该传感器的精度和局限性进行了分析。我们比较了从同时记录的双程图像和哈特曼-夏克图像中获得的眼调制传递函数。评估了影响传感器性能的以下因素:统计精度、用于重建波前的模式数量、微透镜的尺寸和曝光时间。