Suppr超能文献

用于人眼快速全视野周边屈光测量的仪器。

Instrument for fast whole-field peripheral refraction in the human eye.

作者信息

Fernandez Enrique J, Sager Santiago, Lin Zhenghua, Hao Jiangdong, Roca Javier, Prieto Pedro M, Yang Zhikuang, Lan Weizhong, Artal Pablo

机构信息

Laboratorio de Óptica, Centro de Investigación en Óptica y Nanofísica (CiOyN), Universidad de Murcia, Campus de Espinardo (Ed. 34), 30010 Murcia, Spain.

Voptica S.L., Parque Científico de Murcia, Ctra. de Madrid 388, Complejo de Espinardo - Edificio T, 30100 Murcia, Spain.

出版信息

Biomed Opt Express. 2022 Apr 21;13(5):2947-2959. doi: 10.1364/BOE.457686. eCollection 2022 May 1.

Abstract

An instrument for fast and objective measurement of the peripheral refraction in the human eye is presented. The apparatus permits the automatic estimation of both defocus and astigmatism at any retinal eccentricity by scanning a near infrared beam. The design includes a Hartmann-Shack wavefront sensor and a steering mirror, which operate in combination with a compounded eyepiece for wide field operation. The basic scanning protocol allows the estimation of refraction in a circular retinal patch of 50 deg diameter (±25 from central fixation) in 3 sec. Combined with additional fixation points, wider retinal fields can be sampled to achieve a whole field. The instrument underwent calibration and testing, and its performance for real eyes was assessed in 11 subjects of varying age and refraction. The results show high repeatability and precision. The instrument provides a new tool for the investigation of peripheral optics in the human eye.

摘要

本文介绍了一种用于快速、客观测量人眼周边屈光的仪器。该设备通过扫描近红外光束,能够自动估计任何视网膜偏心度下的离焦和散光。其设计包括一个哈特曼-夏克波前传感器和一个转向镜,它们与一个复合目镜配合工作,以实现宽视场操作。基本扫描协议允许在3秒内估计直径为50度(相对于中央注视点±25度)的圆形视网膜区域的屈光情况。结合额外的注视点,可以对更宽的视网膜区域进行采样以实现全视野。该仪器经过了校准和测试,并在11名年龄和屈光状态各异的受试者中评估了其对真实眼睛的性能。结果显示出高重复性和高精度。该仪器为研究人眼周边光学提供了一种新工具。

相似文献

1
Instrument for fast whole-field peripheral refraction in the human eye.
Biomed Opt Express. 2022 Apr 21;13(5):2947-2959. doi: 10.1364/BOE.457686. eCollection 2022 May 1.
2
Central and peripheral refraction measured by a novel double-pass instrument.
Biomed Opt Express. 2023 May 10;14(6):2608-2617. doi: 10.1364/BOE.489881. eCollection 2023 Jun 1.
3
Peripheral Refraction Validity of the Shin-Nippon SRW5000 Autorefractor.
Optom Vis Sci. 2016 Oct;93(10):1254-61. doi: 10.1097/OPX.0000000000000954.
4
In vivo and in vitro repeatability of Hartmann-Shack aberrometry.
J Cataract Refract Surg. 2003 Dec;29(12):2295-301. doi: 10.1016/s0886-3350(03)00655-2.
5
Assessment of objective and subjective eccentric refraction.
Optom Vis Sci. 2005 Apr;82(4):298-306. doi: 10.1097/01.opx.0000159366.61943.62.
6
Wavefront Derived Refraction and Full Eye Biometry in Pseudophakic Eyes.
PLoS One. 2016 Mar 24;11(3):e0152293. doi: 10.1371/journal.pone.0152293. eCollection 2016.
7
Adaptive optics visual simulator.
J Refract Surg. 2002 Sep-Oct;18(5):S634-8. doi: 10.3928/1081-597X-20020901-27.
9
Comparison of two scanning instruments to measure peripheral refraction in the human eye.
J Opt Soc Am A Opt Image Sci Vis. 2012 Mar 1;29(3):258-64. doi: 10.1364/JOSAA.29.000258.
10
Two-dimensional Peripheral Refraction and Retinal Image Quality in Emmetropic Children.
Sci Rep. 2019 Nov 7;9(1):16203. doi: 10.1038/s41598-019-52533-7.

引用本文的文献

1
IMI-Instrumentation for Myopia Management.
Invest Ophthalmol Vis Sci. 2025 Jul 1;66(9):7. doi: 10.1167/iovs.66.9.7.
2
Experimental and modeling analysis of lenses with concentric cylindrical annular refractive elements: impact on peripheral imaging.
Biomed Opt Express. 2025 Mar 7;16(4):1344-1358. doi: 10.1364/BOE.546942. eCollection 2025 Apr 1.
3
Ultra-wide-angle peripheral refraction using a laser-scanning instrument.
Biomed Opt Express. 2024 Oct 24;15(11):6486-6498. doi: 10.1364/BOE.540051. eCollection 2024 Nov 1.
4
Relative peripheral refraction in school children with different refractive errors using a novel multispectral refraction topographer.
Int J Ophthalmol. 2024 Aug 18;17(8):1477-1482. doi: 10.18240/ijo.2024.08.13. eCollection 2024.
5
Widefield wavefront sensor for multidirectional peripheral retinal scanning.
Biomed Opt Express. 2023 Jul 20;14(8):4190-4204. doi: 10.1364/BOE.491412. eCollection 2023 Aug 1.
6
Individualized modeling for the peripheral optics of the human myopic eye.
Biomed Opt Express. 2023 May 17;14(6):2726-2735. doi: 10.1364/BOE.489792. eCollection 2023 Jun 1.

本文引用的文献

1
Perception and Action in Sports. On the Functionality of Foveal and Peripheral Vision.
Front Sports Act Living. 2020 Jan 9;1:66. doi: 10.3389/fspor.2019.00066. eCollection 2019.
2
Two-dimensional peripheral refraction and retinal image quality in orthokeratology lens wearers.
Biomed Opt Express. 2020 Jun 4;11(7):3523-3533. doi: 10.1364/BOE.397077. eCollection 2020 Jul 1.
3
Dual-angle open field wavefront sensor for simultaneous measurements of the central and peripheral human eye.
Biomed Opt Express. 2020 May 18;11(6):3125-3138. doi: 10.1364/BOE.391548. eCollection 2020 Jun 1.
4
Peripheral Vision Tests in Sports: Training Effects and Reliability of Peripheral Perception Test.
Int J Environ Res Public Health. 2019 Dec 9;16(24):5001. doi: 10.3390/ijerph16245001.
5
Two-dimensional Peripheral Refraction and Retinal Image Quality in Emmetropic Children.
Sci Rep. 2019 Nov 7;9(1):16203. doi: 10.1038/s41598-019-52533-7.
6
More than the Useful Field: Considering peripheral vision in driving.
Appl Ergon. 2017 Nov;65:316-325. doi: 10.1016/j.apergo.2017.07.009. Epub 2017 Jul 28.
7
Effects of acute peripheral/central visual field loss on standing balance.
Exp Brain Res. 2017 Nov;235(11):3261-3270. doi: 10.1007/s00221-017-5045-x. Epub 2017 Aug 1.
8
Three-dimensional MRI study of the relationship between eye dimensions, retinal shape and myopia.
Biomed Opt Express. 2017 Apr 5;8(5):2386-2395. doi: 10.1364/BOE.8.002386. eCollection 2017 May 1.
9
Peripheral refraction and higher-order aberrations with cycloplegia and fogging lenses using the BHVI-EyeMapper.
J Optom. 2016 Jan-Mar;9(1):5-12. doi: 10.1016/j.optom.2015.06.003. Epub 2015 Jul 17.
10
Compact optical integration instrument to measure intraocular straylight.
Biomed Opt Express. 2014 Aug 14;5(9):3036-41. doi: 10.1364/BOE.5.003036. eCollection 2014 Sep 1.

文献AI研究员