Detection of bile reflux: in vivo validation of the Bilitec fibreoptic system.

作者信息

Barrett M W, Myers J C, Watson D I, Jamieson G G

机构信息

University of Adelaide Department of Surgery, Royal Adelaide Hospital, South Australia, Australia.

出版信息

Dis Esophagus. 2000;13(1):44-50. doi: 10.1046/j.1442-2050.2000.00062.x.

Abstract

The Bilitec ambulatory bile reflux monitor is advocated for the assessment of bile reflux. However, it has only been validated by bench top (in vitro) studies. This in vivo study used controlled 'iatrogenic' episodes of bile reflux to determine the ability of the Bilitec system to detect bile acids in the oesophageal lumen. At least 10 ml of fluid was aspirated from the stomach of each of 32 patients, and analysed quantitatively for total bile acids (TBA) and total bilirubin (TB). An in vitro assessment of this fluid with the Bilitec probe was performed initially. The Bilitec probe was then sited in the distal oesophagus, and the gastric fluid was introduced into the mid-oesophagus, through a naso-oesophageal tube, as a series of 10-ml volumes of varying dilutions (1:8, 1:4, 1:2, undiluted) to simulate gastro-oesophageal reflux events. Intraoesophageal bilirubin absorbance was compared with TBA and TB for each dilution. Both TBA and TB values correlated strongly with bilirubin absorbance in vitro (r = 0.83, p < 0.0001 and r = 0.82, p < 0.0001 respectively). However, this relationship was weaker in vivo (r = 0.64, p = 0.0001 and r = 0.68, p < 0.0001 respectively). Of the dilutions which contained potentially injurious concentrations of bile acid (>1 mmol/L), 77% were associated with absorbance >0.14 (currently accepted absorbance threshold for oesophageal bile reflux). Although in vitro sensitivity of the Bilitec probe to bile acids was excellent, in vivo sensitivity was less reliable, and 23% of significant 'bile reflux' episodes did not exceed the currently recommended Bilitec absorbance threshold for bile reflux. In contrast, false-positive results were unlikely. These findings suggest that the overall accuracy of the Bilitec system is probably sufficient for clinical use, although this method is not sufficiently reliable to be regarded as the gold standard for the evaluation of duodeno-oesophageal reflux.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验