The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps.

作者信息

Weiss M, Fischer J, Neff T, Baenziger O

机构信息

Department of Intensive Care and Neonatology, University Children's Hospital, Steinwiesstrase 75, CH-8032 Zurich, Switzerland.

出版信息

Anaesthesia. 2000 Nov;55(11):1094-8. doi: 10.1046/j.1365-2044.2000.01550.x.

Abstract

Fluid delivery from four types of commercially available 50-ml syringes was measured using an electronic balance at an infusion rate of 1 ml.h(-1). Retrograde aspiration volume and zero-drug delivery time were recorded after lowering the syringe pump by 50 cm. Syringe compliance was calculated from the volume of bolus released after occlusion at 100 mmHg. Zero-drug delivery times differed significantly between syringes, ranging from [mean (SD)] 3.26 (0.40) min to 6.38 (0.56) min (F = 55.5, d.f. = 3/20, p < 0.0001). Syringe compliance correlated well with aspiration volume (Pearson r(2) = 0.92, p < 0.001) and zero-drug delivery time (r(2) = 0.90, p < 0.001). Syringe design affected the internal syringe compliance. All syringes were associated with potentially relevant zero-drug delivery times after moderate vertical displacement. To minimise this risk, vertical displacement of syringe pumps delivering highly vasoactive drugs should be avoided.

摘要

使用电子天平以1 ml.h⁻¹的输注速率测量了四种市售50毫升注射器的液体输送量。将注射泵降低50厘米后,记录逆行抽吸量和零药物输送时间。注射器顺应性由在100 mmHg压力下阻塞后释放的推注体积计算得出。不同注射器的零药物输送时间差异显著,范围为[平均值(标准差)]3.26(0.40)分钟至6.38(0.56)分钟(F = 55.5,自由度 = 3/20,p < 0.0001)。注射器顺应性与抽吸量(Pearson r² = 0.92,p < 0.001)和零药物输送时间(r² = 0.90,p < 0.001)密切相关。注射器设计影响注射器内部顺应性。在适度垂直位移后,所有注射器都存在潜在的相关零药物输送时间。为了将这种风险降至最低,应避免输送高血管活性药物的注射泵发生垂直位移。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验