玻璃体切除术中的眼内照明及光毒性阈值。

Endoillumination during vitrectomy and phototoxicity thresholds.

作者信息

van den Biesen P R, Berenschot T, Verdaasdonk R M, van Weelden H, van Norren D

机构信息

FC Donders Institute of Ophthalmology, University Hospital Utrecht, Utrecht, Netherlands.

出版信息

Br J Ophthalmol. 2000 Dec;84(12):1372-5. doi: 10.1136/bjo.84.12.1372.

Abstract

AIM

To assess the retinal phototoxicity hazards of and to provide safety margins for endoillumination during vitrectomy.

METHODS

The absolute power and spectral distribution from various light sources and filter combinations that are commercially available for vitreous surgery were measured. The maximal exposure times based on the ICNIRP safety guidelines for photochemical and thermal injury of the aphakic eye were calculated. Additionally, the effect of various measures that reduce the risk of phototoxicity was evaluated.

RESULTS

Measurements of the spectrum and energy indicated that the ICNIRP safety guidelines for photochemical retinal damage are exceeded within 1 minute for nine out of 10 combinations tested. With an additional 475 nm long pass filter, light levels below 10 mW, and a distance from light probe to retina of at least 10 mm, the allowable exposure time can be increased up to 13 minutes. Thermal damage can be anticipated when the light probe touches the retina.

CONCLUSION

Commercially available light sources for endoillumination during vitrectomy are not safe with respect to photochemical retinal damage. Even with maximal precautions macular phototoxic damage remains a factual danger during vitrectomy.

摘要

目的

评估玻璃体切除术中眼内照明的视网膜光毒性危害,并提供安全边际。

方法

测量了各种可用于玻璃体手术的市售光源和滤光片组合的绝对功率和光谱分布。根据国际非电离辐射防护委员会(ICNIRP)关于无晶状体眼光化学和热损伤的安全指南,计算了最大暴露时间。此外,还评估了各种降低光毒性风险措施的效果。

结果

光谱和能量测量表明,在所测试的10种组合中,有9种在1分钟内就超过了ICNIRP关于视网膜光化学损伤的安全指南。使用额外的475nm长波滤光片、低于10mW的光强以及光探头与视网膜至少10mm的距离时,允许的暴露时间可增加至13分钟。当光探头接触视网膜时,可预期会发生热损伤。

结论

就视网膜光化学损伤而言,市售的玻璃体切除术中眼内照明光源不安全。即使采取最大程度的预防措施,黄斑光毒性损伤在玻璃体切除术期间仍然是一个实际存在的危险。

相似文献

1
Endoillumination during vitrectomy and phototoxicity thresholds.
Br J Ophthalmol. 2000 Dec;84(12):1372-5. doi: 10.1136/bjo.84.12.1372.
2
The price for reduced light toxicity: Do endoilluminator spectral filters decrease color contrast during Brilliant Blue G-assisted chromovitrectomy?
Graefes Arch Clin Exp Ophthalmol. 2014 Mar;252(3):367-74. doi: 10.1007/s00417-013-2461-x. Epub 2013 Sep 8.
3
Illumination and phototoxicity issues in vitreoretinal surgery.
Retina. 2008 Jan;28(1):1-4. doi: 10.1097/IAE.0b013e318156e015.
4
[Possible damage to the eye caused by light from ophthalmologic equipment].
Klin Monbl Augenheilkd. 1991 Jun;199(1):62-6. doi: 10.1055/s-2008-1046049.
5
[Retinal phototoxicity from endoilluminators for vitrectomy].
Nippon Ganka Gakkai Zasshi. 2008 Nov;112(11):975-83.
6
[Emission limits and standards regarding endoilluminators].
Ophthalmologe. 2008 Oct;105(10):911-5. doi: 10.1007/s00347-008-1796-x.
7
Phototoxicity to the retina: mechanisms of damage.
Int J Toxicol. 2002 Nov-Dec;21(6):473-90. doi: 10.1080/10915810290169909.
8
Retinal endoilluminator toxicity of xenon and light-emitting diode (LED) light source: rabbit model.
Cutan Ocul Toxicol. 2014 Sep;33(3):192-6. doi: 10.3109/15569527.2013.832282. Epub 2013 Oct 22.
9
Operating microscope-induced retinal phototoxicity during pars plana vitrectomy.
Arch Ophthalmol. 1988 Apr;106(4):521-3. doi: 10.1001/archopht.1988.01060130567038.

引用本文的文献

1
A case of choroidal neovascularization following multiple repairs of a rhegmatogenous retinal detachment.
Saudi J Ophthalmol. 2025 Mar 24;39(1):104-106. doi: 10.4103/sjopt.sjopt_137_23. eCollection 2025 Jan-Mar.
4
Oxidative Stress and Antioxidant-Based Interventional Medicine in Ophthalmology.
Pharmaceuticals (Basel). 2023 Aug 12;16(8):1146. doi: 10.3390/ph16081146.
6
Two-port dry vitrectomy for rhegmatogenous retinal detachment: a pilot study.
Eye (Lond). 2023 Dec;37(18):3801-3806. doi: 10.1038/s41433-023-02617-6. Epub 2023 Jun 10.
7
Comment on: Amniotic membrane for covering high myopic macular hole associated with retinal detachment following failed primary surgery.
Int J Ophthalmol. 2023 Feb 18;16(2):328-332. doi: 10.18240/ijo.2023.02.25. eCollection 2023.
8
Femtosecond laser-cut autologous anterior lens capsule transplantation to treat refractory macular holes.
Eye (Lond). 2023 Apr;37(6):1073-1079. doi: 10.1038/s41433-022-02062-x. Epub 2022 Apr 15.
9
Three-dimensional heads-up surgery in ab-interno trabeculotomy: Image processing-assisted trabeculotomy.
PLoS One. 2022 Feb 10;17(2):e0263588. doi: 10.1371/journal.pone.0263588. eCollection 2022.
10
3D Heads-Up Display vs. Standard Operating Microscope Vitrectomy for Rhegmatogenous Retinal Detachment.
Front Med (Lausanne). 2020 Dec 22;7:615515. doi: 10.3389/fmed.2020.615515. eCollection 2020.

本文引用的文献

1
Long-term follow-up of iatrogenic phototoxicity.
Arch Ophthalmol. 1998 Jun;116(6):753-7. doi: 10.1001/archopht.116.6.753.
2
Vision-threatening complications of surgery for full-thickness macular holes. Vitrectomy for Macular Hole Study Group.
Ophthalmology. 1997 Sep;104(9):1442-52; discussion 1452-3. doi: 10.1016/s0161-6420(97)30118-3.
3
4
Surgical management of idiopathic epiretinal membranes.
Ophthalmology. 1986 Jul;93(7):978-83. doi: 10.1016/s0161-6420(86)33635-2.
5
Macular phototoxicity caused by fiberoptic endoillumination during pars plana vitrectomy.
Am J Ophthalmol. 1992 Sep 15;114(3):287-96. doi: 10.1016/s0002-9394(14)71792-1.
6
Retinal pigment epitheliopathy after macular hole surgery.
Ophthalmology. 1992 Nov;99(11):1671-7. doi: 10.1016/s0161-6420(92)31746-4.
7
Sensitivity of the retina to radiation damage as a function of wavelength.
Photochem Photobiol. 1979 Apr;29(4):735-43. doi: 10.1111/j.1751-1097.1979.tb07759.x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验