Department of Ophthalmology, University of Tokyo Graduate School of Medicine, Tokyo, Japan.
Department of Ophthalmology, Shinseikai Toyama Hospital, Toyama, Japan.
PLoS One. 2022 Feb 10;17(2):e0263588. doi: 10.1371/journal.pone.0263588. eCollection 2022.
We compared the visibility and surgeon posture between image-processing-assisted trabeculotomy (IP-LOT) using the NGENUITY® 3D visual system and conventional microsurgery (microscope-assisted trabeculotomy; MS-LOT).
IP-LOT was performed for five pig eyes. The visibility of the trabecular mesh work was evaluated on images of the trabecular mesh work and the posterior surface of the cornea (Cor) obtained under three different conditions. Images were then analyzed using ImageJ® to measure differences in luminance between the trabecular mesh work and Cor. IP-LOT was also performed for eleven human eyes, and the data were analyzed using the same approach as that used for the pig eyes. The length from the surgeon's abdomen to the operative eye (working distance) during MS-LOT and IP-LOT was measured for 12 different surgeons and compared to evaluate surgeon posture.
Image processing significantly increased the difference in luminance between the trabecular mesh work and Cor in both pig and human eyes (p < 0.05). Moreover, the working distance in IP-LOT was significantly shorter than that in MS-LOT (p < 0.05).
Our findings suggest that the NGENUITY® 3D visual system provides better trabecular mesh work visibility than a normal microscope in conventional surgical methods, and it allows surgeons to operate without moving far from the operative eye.
我们比较了使用 NGENUITY® 3D 视觉系统的图像处理辅助小梁切开术 (IP-LOT) 和传统显微镜辅助小梁切开术 (MS-LOT) 的可视性和手术医生的体位。
对五只猪眼进行了 IP-LOT。在三种不同条件下,对小梁网和角膜后表面 (Cor) 的小梁网图像进行评估,以评估小梁网的可视性。然后使用 ImageJ® 分析图像,以测量小梁网和 Cor 之间的亮度差异。还对 11 只人眼进行了 IP-LOT,并使用与猪眼相同的方法进行数据分析。测量了 12 名不同手术医生进行 MS-LOT 和 IP-LOT 时从手术医生腹部到手术眼的距离 (工作距离),并进行比较以评估手术医生的体位。
图像处理显著增加了猪眼和人眼中小梁网和 Cor 之间的亮度差异 (p < 0.05)。此外,IP-LOT 的工作距离明显短于 MS-LOT (p < 0.05)。
我们的研究结果表明,与传统手术方法中的普通显微镜相比,NGENUITY® 3D 视觉系统可提供更好的小梁网可视性,并且可以让手术医生在不远离手术眼的情况下进行操作。