Soulantzou Konstantina, Plakitsi Athina, Chalkiadaki Evangelia, Anastasopoulou Maria, Karmiris Efthymios
Department of Biomedical Sciences, Division of Optics and Optometry, University of West Attica, Athens, Greece.
Department of Ophthalmology, 251 Hellenic Airforce General Hospital, Athens, Greece.
Int Ophthalmol. 2023 Feb;43(2):363-370. doi: 10.1007/s10792-022-02431-1. Epub 2022 Jul 21.
To compare and analyze the interchangeability in measuring central corneal thickness (CCT) using ultrasonic pachymetry (USP, PACHMATE 2, DGH, Inc, Exton, PA, USA), non-contact specular microscopy (NCSM, CEM-530, Nidek CO, LTD, Gamagori, Japan) and a high-resolution Scheimpflug Camera (Pentacam HR, OCULUS, Wetzlar, Germany).
An observational, cross-sectional study was performed recruiting 216 volunteers, for a sample size of 216 eyes with no ocular abnormalities other than refractive errors. All subjects underwent pachymetric measurements obtained by USP, NCSM and Pentacam HR. Examinations were performed by the same examiner with USP always following the noncontact examinations.
The mean CCT (± SD) was 560.30 ± 38.80 μm, 556.76 ± 36.83 μm and 547.31 ± 35.28 μm for USP, NCSM and Pentacam HR, respectively. The Bland-Altman analysis showed that the highest concordance was found between USP and NCSM, with differences between - 13.18 μm and 20.26 μm. For the differences between measurements obtained with Pentacam HR and USP, the differences at Bland-Altman plot were between - 28.25 and 13.57 μm. The lowest concordance was found for the CCT values measured with Pentacam HR and NCSM, with differences between - 25.67 and 6.86 μm. The intraclass correlation coefficient (ICC) between all pairs of measurements was between 0.979 and 0.987, suggesting that the mean measurements were strongly correlated.
Measurements obtained with all three devices had high correlation. USP and NCSM were found in good agreement and high concordance, too. The above results indicate that these two devices are interchangeable in clinical practice. Pentacam HR may be a useful alternative for measuring CCT; however, it significantly underestimates CCT and cannot be used interchangeably with the other devices that we studied.
比较并分析使用超声测厚法(USP,PACHMATE 2,DGH公司,美国宾夕法尼亚州埃克斯顿)、非接触式角膜内皮显微镜检查法(NCSM,CEM - 530,日本尼德克有限公司,蒲郡)和高分辨率眼前节分析系统(Pentacam HR,德国韦茨拉尔OCULUS公司)测量中央角膜厚度(CCT)的互换性。
进行一项观察性横断面研究,招募216名志愿者,样本量为216只眼,除屈光不正外无其他眼部异常。所有受试者均接受了由USP、NCSM和Pentacam HR进行的测厚测量。检查由同一名检查者进行,使用USP时总是在非接触检查之后。
USP、NCSM和Pentacam HR测量的平均CCT(±标准差)分别为560.30±38.80μm、556.76±36.83μm和547.31±35.28μm。Bland - Altman分析表明,USP和NCSM之间的一致性最高,差异在 - 13.18μm至20.26μm之间。对于Pentacam HR和USP测量结果之间的差异,Bland - Altman图上的差异在 - 28.25至13.57μm之间。Pentacam HR和NCSM测量的CCT值一致性最低,差异在 - 25.67至6.86μm之间。所有测量对之间的组内相关系数(ICC)在0.979至0.987之间,表明平均测量值高度相关。
使用这三种设备获得的测量结果具有高度相关性。发现USP和NCSM也具有良好的一致性和高度的协调性。上述结果表明这两种设备在临床实践中可互换使用。Pentacam HR可能是测量CCT的一种有用替代方法;然而,它显著低估了CCT,不能与我们研究的其他设备互换使用。