Abstract

PURPOSE

The purpose of this study was to compare and evaluate the agreement of central corneal thickness (CCT) values obtained with three different devices working according to optical principle in healthy eyes.

METHODS

60 eyes of 60 individuals (30 men and 30 women) were enrolled in this study. CCT measurements performed with Scheimpflug-Placido topographer (Sirius), spectral-domain optical coherence tomography (RTVue) with an anterior segment module, and optical biometer (AL-Scan) were compared. Bland-Altman analysis was used to demonstrate agreement between methods.

RESULTS

The mean age was 30.07 ± 7.313 years (range, 18-47 years). The mean CCT values obtained by RTVue, Sirius, and AL-Scan were 518.25 ± 36.38 µm, 526.08 ± 36.33 µm, and 513.50 ± 39.09 µm, respectively. The mean differences in CCT were 7.83 ± 14.15 µm between Sirius and RTVue, 12.58 ± 11.87 µm between Sirius and AL-Scan, and 4.75 ± 4.50 µm between RTVue and AL-Scan. The mean CCT was statistically different among the three groups ( < 0.05). All three modalities of CCT measurements correlated closely with each other, with Pearson's correlation coefficients ranging from 0.924 to 0.961. The 95% limits of agreement were -19.90 to 35.56 µm between Sirius and RTVue, -10.69 to 35.85 µm between Sirius and AL-Scan, and -4.07 to 13.58 µm between RTVue and AL-Scan.

CONCLUSION

Different results could be obtained through different noncontact devices in CCT measurements. Although the measurement values obtained by these devices show a high level of correlation, it would be a more correct approach to not use them directly interchangeably in clinical practice. Evaluation and follow-up of CCT should be performed using the same device.

摘要

目的

本研究的目的是比较和评估在健康眼睛中使用三种基于光学原理工作的不同设备获得的中央角膜厚度(CCT)值的一致性。

方法

本研究纳入了60名个体(30名男性和30名女性)的60只眼睛。比较了使用Scheimpflug - Placido地形图仪(Sirius)、带前段模块的光谱域光学相干断层扫描(RTVue)和光学生物测量仪(AL - Scan)进行的CCT测量。采用Bland - Altman分析来证明各方法之间的一致性。

结果

平均年龄为30.07±7.313岁(范围18 - 47岁)。RTVue、Sirius和AL - Scan获得的平均CCT值分别为518.25±36.38µm、526.08±36.33µm和513.50±39.09µm。Sirius与RTVue之间CCT的平均差异为7.83±14.15µm,Sirius与AL - Scan之间为12.58±11.87µm,RTVue与AL - Scan之间为4.75±4.50µm。三组之间的平均CCT在统计学上存在差异(<0.05)。所有三种CCT测量方式彼此密切相关,Pearson相关系数范围为0.924至0.961。Sirius与RTVue之间的95%一致性界限为 - 19.90至35.56µm,Sirius与AL - Scan之间为 - 10.69至35.85µm,RTVue与AL - Scan之间为 - 4.07至13.58µm。

结论

在CCT测量中,不同的非接触设备可能会得到不同的结果。尽管这些设备获得的测量值显示出高度相关性,但在临床实践中不直接将它们互换使用会是更正确的做法。CCT的评估和随访应使用同一设备进行。