采用氦光离子化脉冲放电检测的静态顶空气相色谱法同时测定矿物绝缘油中的溶解气体和水分。

Simultaneous determination of dissolved gases and moisture in mineral insulating oils by static headspace gas chromatography with helium photoionization pulsed discharge detection.

作者信息

Jalbert J, Gilbert R, Tétreault P

机构信息

Institut de recherche d'Hydro-Québec, Varennes, Québec, Canada.

出版信息

Anal Chem. 2001 Jul 15;73(14):3382-91. doi: 10.1021/ac010063f.

Abstract

This paper presents the development of a static headspace capillary gas chromatographic method (HS-GC) for simultaneously determining dissolved gases (H2, O2, N2, CO, CO2, CH4, C2H6, C2H4, C2H2, C3H8) and moisture from a unique 15-mL mineral oil sample. A headspace sampler device is used to equilibrate the sample species in a two-phase system under controlled temperature and agitation conditions. A portion of the equilibrated species is then automatically split-injected into two chromatographic channels mounted on the same GC for their separation. The hydrocarbons and the lighter gases are separated on the first channel by a GS-Q column coupled with a MolSieve 5-A column via a bypass valve, while the moisture is separated on the second channel using a Stabilwax column. The analytes are detected by using two universal pulsed-discharge helium ionization detectors (PDHID). The performance of the method was established using equilibrated vials containing known amounts of gas mixture, water, and blank oil. The signal is linear over the concentration ranges normally found for samples collected from open-breathing power transformers. Determination sensitivity varies with the nature of the species considered with values as high as 21 500 A x 10(-9) s (microg/ g)(-1) for H2O, 46-216 A x 10(-9) s (microL/L)(-1) for the hydrocarbons and carbon oxides, and as low as 8-21 A x 10(-9) s (microL/L)(-1) for the O2 and N2 permanent gases. The detection limit of the method is between 0.08 and 6 microL/L for the dissolved gases, except for O2, N2, and CO2, where higher values are observed due to air intrusion during sampler operations, and 0.1 microg/g for the dissolved water. Ten consecutive measurements in the low and high levels of the calibration curves have shown a precision better than 12% and 6%, respectively, in all cases. A comparison study between the HS-GC method and the ASTM standard procedures on 31 field samples showed a very good agreement of the results. The advantages of configuring the arrangement with two PDHID over the conventional flame ionization and thermal conductivity detectors were clearly demonstrated.

摘要

本文介绍了一种静态顶空毛细管气相色谱法(HS-GC)的开发,该方法用于同时测定取自独特的15 mL矿物油样品中的溶解气体(H2、O2、N2、CO、CO2、CH4、C2H6、C2H4、C2H2、C3H8)和水分。使用顶空进样器装置在受控的温度和搅拌条件下使样品中的物质在两相系统中达到平衡。然后将一部分平衡后的物质自动分流进样到安装在同一台气相色谱仪上的两个色谱通道中进行分离。烃类和较轻的气体在第一个通道上通过一根GS-Q柱与一根5A分子筛柱通过一个旁路阀相连进行分离,而水分在第二个通道上使用一根Stabilwax柱进行分离。使用两个通用脉冲放电氦离子化检测器(PDHID)对分析物进行检测。该方法的性能通过使用含有已知量气体混合物、水和空白油的平衡瓶来确定。在从开放式呼吸电力变压器采集样品通常遇到的浓度范围内,信号呈线性。测定灵敏度随所考虑物质的性质而变化,对于H2O高达21500 A×10(-9) s (μg/g)(-1),对于烃类和碳氧化物为46 - 216 A×10(-9) s (μL/L)(-1),对于O2和N2等永久性气体低至8 - 21 A×10(-9) s (μL/L)(-1)。该方法对溶解气体的检测限除O2、N2和CO2外为0.08至6 μL/L,由于进样器操作过程中空气侵入,O2、N2和CO2的检测限较高,对溶解水的检测限为0.1 μg/g。在校准曲线的低水平和高水平下进行的十次连续测量表明,在所有情况下精密度分别优于12%和6%。对31个现场样品进行的HS-GC方法与ASTM标准程序之间的比较研究表明,结果非常吻合。与传统的火焰离子化检测器和热导检测器相比,配置两个PDHID的优势得到了明显体现。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验