Holmes J M, Beck R W, Repka M X, Leske D A, Kraker R T, Blair R C, Moke P S, Birch E E, Saunders R A, Hertle R W, Quinn G E, Simons K A, Miller J M
Mayo Clinic, Ophthalmology West 7, Rochester, MN 55905, USA.
Arch Ophthalmol. 2001 Sep;119(9):1345-53. doi: 10.1001/archopht.119.9.1345.
To evaluate the reliability of a new visual acuity testing protocol for children using isolated surrounded HOTV optotypes.
After initial pilot testing and modification, the protocol was evaluated using the Baylor-Video Acuity Tester (BVAT) to present isolated surrounded HOTV optotypes. At 6 sites, the protocol was evaluated for testability in 178 children aged 2 to 7 years and for reliability in a subset of 88 children. Twenty-eight percent of the 178 children were classified as having amblyopia.
Using the modified protocol, testability ranged from 24% in 2-year-olds to 96% in 5- to 7-year-olds. Test-retest reliability was high (r = 0.82), with 93% of retest scores within 0.1 logMAR unit of the initial test score. The 95% confidence interval for an acuity score was calculated to be the score +/-0.125 logMAR unit. For a change between 2 acuity scores, the 95% confidence interval was the difference +/-0.18 logMAR unit.
The visual acuity protocol had a high level of testability in 3- to 7-year-olds and excellent test-retest reliability. The protocol has been incorporated into the multicenter Amblyopia Treatment Study and has wide potential application for standardizing visual acuity testing in children.
评估一种使用孤立的周边HOTV视标对儿童进行视力测试的新方案的可靠性。
在初步试点测试和修改后,使用贝勒视频视力测试仪(BVAT)来呈现孤立的周边HOTV视标,对该方案进行评估。在6个地点,对178名2至7岁儿童的该方案进行了可测试性评估,并对其中88名儿童的子集进行了可靠性评估。178名儿童中有28%被归类为弱视。
使用修改后的方案,可测试性范围从2岁儿童的24%到5至7岁儿童的96%。重测信度较高(r = 0.82),93%的重测分数在初始测试分数的0.1 logMAR单位范围内。视力分数的95%置信区间计算为分数±0.125 logMAR单位。对于两个视力分数之间的变化,95%置信区间为差异±0.18 logMAR单位。
该视力测试方案在3至7岁儿童中具有较高的可测试性和出色的重测信度。该方案已被纳入多中心弱视治疗研究,在标准化儿童视力测试方面具有广泛的潜在应用。