在正常、炎热潮湿条件下对乳酸Pro分析仪、Accusport分析仪、Analox GM7分析仪和柯达Ektachem乳酸分析仪的比较。

A comparison of the lactate Pro, Accusport, Analox GM7 and Kodak Ektachem lactate analysers in normal, hot and humid conditions.

作者信息

Mc Naughton L R, Thompson D, Philips G, Backx K, Crickmore L

机构信息

University of Bath, Claverton Down, Bath BA 2 7AY, Great Britain, UK.

出版信息

Int J Sports Med. 2002 Feb;23(2):130-5. doi: 10.1055/s-2002-20133.

Abstract

This study aimed to compare the performance of a new portable lactate analyser against other standard laboratory methods in three conditions, normal (20 +/- 1.3 degrees C; 40 +/- 5 % RH), hot (40 +/- 2.5 degrees C; 40 +/- 5 % RH), and humid (20 +/- 1.1 degrees C; 82 +/- 6 % RH) conditions. Seven healthy males, ([Mean +/- SE]: age, 26.3 +/- 1.3 yr; height, 177.7 +/- 1.6 cm; weight, 77.4 +/- 0.9 kg, .VO(2)max, 56.1 +/- 1.9 ml x kg x min(-1)) undertook a maximal cycle ergometry test to exhaustion in the three conditions. Blood was taken every 3 min at the end of each stage and was analysed using the Lactate Pro LT-1710, the Accusport, the Analox GM7 and the Kodak Ektachem systems. The MANOVA (Analyser Type x Condition x Workload) indicated no interaction effect (F(42,660), = 0.45, p > 0.99, Power = 0.53). The data across all workloads indicated that the machines measured significantly differently to each other (F(4,743) = 14.652, p < 0.0001, Power = 1.00). The data were moderately to highly correlated. We conclude that the Lactate Pro is a simple and effective measurement device for taking blood lactate in a field or laboratory setting. However, we would caution against using this machine to compare data from other machines.

摘要

本研究旨在比较一种新型便携式乳酸分析仪与其他标准实验室方法在三种条件下的性能,这三种条件分别为正常条件(20±1.3℃;40±5%相对湿度)、炎热条件(40±2.5℃;40±5%相对湿度)和潮湿条件(20±1.1℃;82±6%相对湿度)。七名健康男性([平均值±标准误]:年龄,26.3±1.3岁;身高,177.7±1.6厘米;体重,77.4±0.9千克,最大摄氧量,56.1±1.9毫升·千克·分钟⁻¹)在这三种条件下进行了最大运动强度的自行车测力计测试直至力竭。在每个阶段结束时每隔3分钟采集一次血液,并使用乳酸Pro LT - 1710、Accusport、Analox GM7和柯达Ektachem系统进行分析。多变量方差分析(分析仪类型×条件×工作量)表明不存在交互作用效应(F(42,660) = 0.45,p>0.99,检验效能 = 0.53)。所有工作量的数据表明,各台机器的测量结果彼此之间存在显著差异(F(4,743) = 14.652,p<0.0001,检验效能 = 1.00)。数据具有中度到高度的相关性。我们得出结论,乳酸Pro是一种在现场或实验室环境中采集血乳酸的简单有效的测量设备。然而,我们提醒不要使用该机器来比较来自其他机器的数据。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验