Ascenti G, Zimbaro G, Mazziotti S, Gaeta M, Settineri N, Scribano E
Institute of Radiological Sciences, Policlinico G. Martino, University of Messina, Italy.
Abdom Imaging. 2001 Nov-Dec;26(6):654-60. doi: 10.1007/s00261-001-0025-8.
In a prospective study, we compared power Doppler with and without contrast medium in the depiction of vascularity for the characterization of hyperechoic renal lesions.
Forty-one hyperechoic renal expansive lesions (29 benign, 12 malignant) in 32 patients were studied with power-Doppler ultrasonography before and after administration of an echo-enhancing agent (Levovist Schering AG, Berlin, Germany). Vascular architecture of the lesions was categorized into five different patterns.
Power Doppler ultrasonography showed vascular structures in 25 lesions. The study enhanced with Levovist showed vascularity in eight of 16 lesions not seen on the unenhanced study. The characterization of vascular patterns with unenhanced power Doppler ultrasonography improved diagnostic accuracy compared with gray-scale ultrasonography (59% vs. 32%). The combination of B mode and power Doppler produced even greater diagnostic accuracy (78%), independent of the administration of echo-enhancing agent. Levovist administration was useful in the differential diagnosis between pseudotumor and neoplasm.
The use of sonographic contrast agent did not increase the diagnostic accuracy of power Doppler in the differential diagnosis of hyperechoic renal lesions but was advantageous for the characterization of suspected pseudomasses.
在一项前瞻性研究中,我们比较了使用和不使用造影剂的能量多普勒成像在显示高回声肾病变血管方面的情况,以对这些病变进行特征性描述。
对32例患者的41个高回声肾实性病变(29个良性,12个恶性)在使用回声增强剂(德国柏林先灵公司的Levovist)前后进行了能量多普勒超声检查。病变的血管结构分为五种不同类型。
能量多普勒超声在25个病变中显示出血管结构。使用Levovist增强后的研究显示,在16个未增强检查时未见血管的病变中有8个显示出血管。与灰阶超声相比,未增强的能量多普勒超声对血管类型的特征性描述提高了诊断准确性(59%对32%)。B模式和能量多普勒的联合使用产生了更高的诊断准确性(78%),且与是否使用回声增强剂无关。Levovist的使用有助于假性肿瘤和肿瘤的鉴别诊断。
超声造影剂的使用并未提高能量多普勒在高回声肾病变鉴别诊断中的诊断准确性,但对疑似假肿块的特征性描述具有优势。