Stone Mary F, Zhao Wei, Jacak Barbara V, O'Connor Paul, Yu Bo, Rehak Pavel
Department of Physics and Astronomy, State University of New York at Stony Brook, 11794-3800, USA.
Med Phys. 2002 Mar;29(3):319-24. doi: 10.1118/1.1449874.
A study of the x-ray sensitivity of amorphous selenium (a-Se) for digital mammography has been performed. A uniform layer of a-Se was deposited on a glass substrate with electrodes on both surfaces. The deposition procedure was identical to that used for a-Se flat-panel detectors. A high voltage was applied to the top surface of the a-Se layer in order to establish an electric field E(Se). Then the sample was exposed to x rays with 27 kVp spectra generated from an x-ray tube with a molybdenum (Mo) target. The mean x-ray energy of the spectrum used was approximately 16.6 keV. The x-ray current generated by the a-Se layer was measured as a function of E(Se). From the current measurement and the estimation of total x-ray energy absorbed in the a-Se, the energy required to create one electron-hole pair (EHP), W, was determined as a function of E(Se). It was found that at the most commonly used E(Se) of 10 V/microm, W was measured as 64 eV. This is considerably higher than the widely accepted typical value of W = 50 eV measured at higher x-ray photon energies (e.g., 50 keV). The dependence of W as a function of E(Se) can be best fitted using the empirical expression of E(Se)-gamma. This relationship is consistent with the results obtained at higher x-ray energies. This article provides an accurate measurement of x-ray sensitivity of a-Se at mammographic energies independent of detector operation, such as the most recently developed flat-panel detectors. The results will be a useful tool for investigation and optimization of a-Se-based x-ray imaging detectors, such as determination of pixel fill-factor and optimal E(Se) during operation.
已对用于数字乳腺摄影的非晶硒(a-Se)的X射线敏感性进行了一项研究。在玻璃基板上沉积了一层均匀的a-Se,其两面都有电极。沉积过程与用于a-Se平板探测器的过程相同。向a-Se层的顶面施加高电压,以建立电场E(Se)。然后,将样品暴露于由带有钼(Mo)靶的X射线管产生的27 kVp光谱的X射线下。所使用光谱的平均X射线能量约为16.6 keV。测量a-Se层产生的X射线电流作为E(Se)的函数。根据电流测量以及对a-Se中吸收的总X射线能量的估计,确定产生一个电子-空穴对(EHP)所需的能量W作为E(Se)的函数。结果发现,在最常用的E(Se)为10 V/μm时,测得W为64 eV。这大大高于在较高X射线光子能量(例如50 keV)下测得的被广泛接受的典型值W = 50 eV。W作为E(Se)的函数的依赖性可以使用E(Se)-γ的经验表达式进行最佳拟合。这种关系与在较高X射线能量下获得的结果一致。本文提供了在乳腺摄影能量下对a-Se的X射线敏感性的精确测量,该测量与探测器操作无关,例如最新开发的平板探测器。这些结果将成为研究和优化基于a-Se的X射线成像探测器的有用工具,例如在操作过程中确定像素填充因子和最佳E(Se)。