用于乳腺摄影的非晶硒的X射线敏感性。

The x-ray sensitivity of amorphous selenium for mammography.

作者信息

Stone Mary F, Zhao Wei, Jacak Barbara V, O'Connor Paul, Yu Bo, Rehak Pavel

机构信息

Department of Physics and Astronomy, State University of New York at Stony Brook, 11794-3800, USA.

出版信息

Med Phys. 2002 Mar;29(3):319-24. doi: 10.1118/1.1449874.

Abstract

A study of the x-ray sensitivity of amorphous selenium (a-Se) for digital mammography has been performed. A uniform layer of a-Se was deposited on a glass substrate with electrodes on both surfaces. The deposition procedure was identical to that used for a-Se flat-panel detectors. A high voltage was applied to the top surface of the a-Se layer in order to establish an electric field E(Se). Then the sample was exposed to x rays with 27 kVp spectra generated from an x-ray tube with a molybdenum (Mo) target. The mean x-ray energy of the spectrum used was approximately 16.6 keV. The x-ray current generated by the a-Se layer was measured as a function of E(Se). From the current measurement and the estimation of total x-ray energy absorbed in the a-Se, the energy required to create one electron-hole pair (EHP), W, was determined as a function of E(Se). It was found that at the most commonly used E(Se) of 10 V/microm, W was measured as 64 eV. This is considerably higher than the widely accepted typical value of W = 50 eV measured at higher x-ray photon energies (e.g., 50 keV). The dependence of W as a function of E(Se) can be best fitted using the empirical expression of E(Se)-gamma. This relationship is consistent with the results obtained at higher x-ray energies. This article provides an accurate measurement of x-ray sensitivity of a-Se at mammographic energies independent of detector operation, such as the most recently developed flat-panel detectors. The results will be a useful tool for investigation and optimization of a-Se-based x-ray imaging detectors, such as determination of pixel fill-factor and optimal E(Se) during operation.

摘要

已对用于数字乳腺摄影的非晶硒(a-Se)的X射线敏感性进行了一项研究。在玻璃基板上沉积了一层均匀的a-Se,其两面都有电极。沉积过程与用于a-Se平板探测器的过程相同。向a-Se层的顶面施加高电压,以建立电场E(Se)。然后,将样品暴露于由带有钼(Mo)靶的X射线管产生的27 kVp光谱的X射线下。所使用光谱的平均X射线能量约为16.6 keV。测量a-Se层产生的X射线电流作为E(Se)的函数。根据电流测量以及对a-Se中吸收的总X射线能量的估计,确定产生一个电子-空穴对(EHP)所需的能量W作为E(Se)的函数。结果发现,在最常用的E(Se)为10 V/μm时,测得W为64 eV。这大大高于在较高X射线光子能量(例如50 keV)下测得的被广泛接受的典型值W = 50 eV。W作为E(Se)的函数的依赖性可以使用E(Se)-γ的经验表达式进行最佳拟合。这种关系与在较高X射线能量下获得的结果一致。本文提供了在乳腺摄影能量下对a-Se的X射线敏感性的精确测量,该测量与探测器操作无关,例如最新开发的平板探测器。这些结果将成为研究和优化基于a-Se的X射线成像探测器的有用工具,例如在操作过程中确定像素填充因子和最佳E(Se)。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验