Noel A, Thibault F
Centre Alexis Vautrin, Unité de Radiophysique Médicale, Paris, France.
Eur Radiol. 2004 Nov;14(11):1990-8. doi: 10.1007/s00330-004-2446-6. Epub 2004 Oct 8.
This paper reviews the different techniques available and competing for full-field digital mammography. The detectors are described in their principles: photostimulable storage phosphor plates inserted as a cassette in a conventional mammography unit, dedicated active matrix detectors (i.e., flat-panel, thin-film transistor-based detectors) and scanning systems, using indirect and direct X-ray conversion. The main parameters that characterize the performances of the current systems and influence the quality of digital images are briefly explained: spatial resolution, detective quantum efficiency and modulation transfer function. Overall performances are often the result of compromises in the choice of technology.
本文回顾了用于全场数字化乳腺摄影的不同可用技术及其竞争情况。文中描述了探测器的原理:作为暗盒插入传统乳腺摄影设备中的光激励存储磷光板、专用有源矩阵探测器(即平板、基于薄膜晶体管的探测器)以及使用间接和直接X射线转换的扫描系统。简要解释了表征当前系统性能并影响数字图像质量的主要参数:空间分辨率、探测量子效率和调制传递函数。总体性能往往是技术选择中折衷的结果。