Three-dimensional CT with a modified C-arm image intensifier: feasibility.

作者信息

Linsenmaier Ulrich, Rock Clemens, Euler Ekkehard, Wirth Stefan, Brandl Roland, Kotsianos Dorothea, Mutschler Wolf, Pfeifer Klaus Jürgen

机构信息

Department of Radiology, Klinikum der Universität München, Innenstadt, Ludwig-Maximilians-Universität München, Nussbaumstrasse 20, D-80336 Munich, Germany.

出版信息

Radiology. 2002 Jul;224(1):286-92. doi: 10.1148/radiol.2241010646.

Abstract

A portable C arm was modified for cone-beam computed tomography (CT). This three-dimensional (3D) CT imaging system facilitated the acquisition of fluoroscopic images during a 190 degrees rotation and computed a 3D data cube (matrix, 256 x 256 x 256; scanning time, 100 seconds) with multiplanar image reformation. The high-contrast resolution, 0.9 line pairs per millimeter, was comparable; the low-contrast resolution, minimal; and the radiation dose, 60%-80% lower, as compared with these parameters at spiral CT. The normal anatomy of small joints could be depicted, and the osteosynthesis screws in the talus were correctly identified.

摘要

一台便携式C形臂被改装用于锥形束计算机断层扫描(CT)。这个三维(3D)CT成像系统在190度旋转过程中便于获取荧光透视图像,并通过多平面图像重建计算出一个3D数据立方体(矩阵,256×256×256;扫描时间,100秒)。与螺旋CT的这些参数相比,其高对比度分辨率为每毫米0.9线对,相当;低对比度分辨率,最低;辐射剂量,低60%-80%。小关节的正常解剖结构可以被描绘出来,距骨中的接骨螺钉也能被正确识别。