AIM

To determine the influence of various irrigating solutions on the accuracy of the electronic apex locator facility in the Tri Auto ZX handpiece.

METHODOLOGY

One hundred and forty teeth with single canals and mature apices, scheduled for extraction for either periodontal or prosthetic reasons in 76 patients were used. Following informed written consent local anaesthesia was administered, access cavities were prepared and pulp tissue removed. The teeth were then randomly divided into seven groups according to the irrigating solutions used. The root canal length measurements were completed using the Tri Auto ZX handpiece with automatic reverse function in the presence of one or other of the following solutions: 0.9% saline, 2.5% NaOCl, 3% H(2)O(2), 0.2% chlorhexidine, 17% EDTA, Ultracaine D-S or in the absence of an irrigating solution (control). Files were immobilized in the access cavity with composite resin. After extraction, the apical regions of the teeth were exposed and the file tips examined under a stereomicroscope. Distances between the file tips and the apical constriction were measured (mm) and analysed using a one-way anova and post hoc Tukey test.

RESULTS

Mean distances from the apical constriction to the file tip were longer in the 0.9% saline group (P<0.05). There was no statistically significant difference on file tip position between the other solutions.

CONCLUSIONS

Tri Auto ZX gave reliable results with all irrigating solutions apart from in the presence of 0.9% saline.

摘要

目的

确定不同冲洗液对Tri Auto ZX手机中电子根尖定位仪准确性的影响。

方法

使用140颗单根管且根尖成熟的牙齿,这些牙齿因牙周或修复原因计划在76例患者中拔除。在获得知情书面同意后,进行局部麻醉,制备开髓洞型并去除牙髓组织。然后根据所使用的冲洗液将牙齿随机分为七组。在以下一种或另一种冲洗液存在的情况下,使用具有自动反转功能的Tri Auto ZX手机完成根管长度测量:0.9%生理盐水、2.5%次氯酸钠、3%过氧化氢、0.2%氯己定、17%乙二胺四乙酸、丙胺卡因D-S或不使用冲洗液(对照组)。锉针用复合树脂固定在开髓洞中。拔牙后,暴露牙齿的根尖区域,并在体视显微镜下检查锉针尖端。测量锉针尖端与根尖狭窄处之间的距离(毫米),并使用单因素方差分析和事后Tukey检验进行分析。

结果

0.9%生理盐水组中从根尖狭窄处到锉针尖端的平均距离更长(P<0.05)。其他冲洗液之间在锉针尖端位置上没有统计学上的显著差异。

结论

除了在0.9%生理盐水存在的情况下,Tri Auto ZX在所有冲洗液中都能给出可靠的结果。