对海德堡视网膜血流计进行改良,以记录图形和闪烁诱发的血流变化。

Modification of the Heidelberg retinal flowmeter to record pattern and flicker induced blood flow changes.

作者信息

Schmeisser Elmar T, Harrison Joseph M, Sutter Erich E, Kiel Jeffrey, Elliott W Rowe, Sponsel W E

机构信息

University of Texas Health Science Center, Department of Ophthalmology, San Antonio, TX, USA.

出版信息

Doc Ophthalmol. 2003 May;106(3):257-63. doi: 10.1023/a:1022973315359.

Abstract

PURPOSE

We tested a prototype stimulator interfaced with a commercially available scanning laser ophthalmoscope designed to measure retinal capillary perfusion (Heidelberg Retina Flowmeter (HRF)). The add-on stimulator optically superimposed the image of a monitor display on to the subject's retina coaxially with the imaging optics of the HRF. The purpose of the study was to determine if flicker and pattern stimulation presented in this manner could evoke changes in retinal perfusion that could be measured by the HRF.

METHODS

The prototype stimulator projected 55 degrees visual angle circular fields of homogeneous flicker, alternating checkerboard, and multi-focal m-sequence hexagonal patterns on the retina of 10 human subjects during acquisition of images by the HRF.

RESULTS

Images were successfully acquired and processed. HRF perfusion values during flicker and pattern stimulation were not significantly different from control values.

CONCLUSIONS

Results of the present study and a previously published study showing a flicker-induced increase in the HRF perfusion values are contradictory. Retinal perfusion measured by the HRF were not affected by flicker and pattern stimulation delivered through the prototype device. These data are not consistent with a large flicker or pattern induced increase in retinal perfusion. The instrumental modification appears promising. However, the raster scan stimulation technique or some other aspect of stimulation or image acquisition may account for the different results in the present study and previous studies in our laboratory and in the laboratories of other investigators.

摘要

目的

我们测试了一种与市售扫描激光检眼镜接口的原型刺激器,该检眼镜旨在测量视网膜毛细血管灌注(海德堡视网膜血流仪(HRF))。附加刺激器将监视器显示屏的图像与HRF的成像光学系统同轴地光学叠加在受试者的视网膜上。本研究的目的是确定以这种方式呈现的闪烁和图案刺激是否能引起可由HRF测量的视网膜灌注变化。

方法

在HRF采集图像期间,原型刺激器在10名人类受试者的视网膜上投射55度视角的均匀闪烁、交替棋盘格和多焦点m序列六边形图案的圆形区域。

结果

成功采集并处理了图像。闪烁和图案刺激期间的HRF灌注值与对照值无显著差异。

结论

本研究结果与先前发表的一项显示闪烁诱导HRF灌注值增加的研究结果相互矛盾。通过原型设备进行的闪烁和图案刺激并未影响HRF测量的视网膜灌注。这些数据与大量闪烁或图案诱导的视网膜灌注增加不一致。仪器的改进似乎很有前景。然而,光栅扫描刺激技术或刺激或图像采集的其他方面可能是本研究与我们实验室以及其他研究者实验室先前研究结果不同的原因。